Lipoprotein(a) analytical variation and cardiovascular thresholds.
Anne-Sophie Bargnoux, Amal Aamoum, Pierre-Edouard Grillet, Maëlle Plawecki, Anne-Marie Dupuy, François Roubille, Jean-Paul Cristol
Département de Biochimie et Hormonologie, CHU de Montpellier, Université de Montpellier, Montpellier, France. · Department of Cardiology, INI-CRT, PhyMedExp, INSERM, CNRS, CHU de Montpellier, Université de Montpellier, Montpellier, France.
Key points
From the paper's conclusion- The percent agreement between two assays varied from 87.6 to 96.6 % and from 82.8 to 93.8 % using the EAS or the ACC/AHA thresholds.
Abstract
OBJECTIVES: The study was designed to evaluate the analytical performances of six automated immunoturbidimetric lipoprotein(a) (Lp(a)) assays, each corresponding to a combination of a reagent and an analyzer (Binding Site Optilite Lp(a) kit, Diagam Lp(a) reagent on Mindray BS480, Diagam Lp(a) reagent on Roche c502, Randox Lp(a) reagent on IDS-ISYS, Roche Tinaquant LPA2 on Roche c702 and Siemens Atellica CH Lp(a)/Atellica® CH 930) using results expressed in mg/dL or nmol/L. METHODS: Imprecision was assessed on manufacturer controls according to the CLSI EP15-A3 protocol. One hundred and 45 patient serum samples from the cardiology department were assessed to compare each method using Spearman's correlation, Passing-Bablok regression and Bland-Altman analyses. Percent agreement was calculated according to the European Atherosclerosis Society and American College of Cardiology/American Heart Association Lp(a) cut-offs for cardiovascular risk. RESULTS: All immunoturbidimetric methods achieved the desirable imprecision of 5.1 % based on biological variation. Highly significant correlations were found across multiple comparisons (p<0.001). The absolute mean of the difference according to Bland-Altman ranged between 0.4 and 2.6 for Lp(a) assays in mg/dL and between 0.6 and 6.5 for Lp(a) assays in nmol/L. The percent agreement between two assays varied from 87.6 to 96.6 % and from 82.8 to 93.8 % using the EAS or the ACC/AHA thresholds. CONCLUSIONS: Despite acceptable correlation among the six Lp(a) turbidimetric assays, a risk of "misclassifications" between two assays up to 12.4 % with the EAS recommendations and 17.2 % with the ACC/AHA guideline remains a major concern. Standardization of methods, including expression of results in nmol/L is needed.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-017 Identifying and Resolving Outliers in Lipoprotein(a) Immunoassay TestingClinical Chemistry · 2026
- Harmonization of Lipoprotein(a) Immunoassays with the IFCC- Endorsed Reference Measurement Procedure Validated in Patient Samples and Multiple Routine testing as A Second Step Towards Apolipoprotein Standardization.Journal of atherosclerosis and thrombosis · 2026
- Implementation of an Automated Lipoprotein(a) Assay on a Third-Party Automated Analyzer: Validation, Method Comparison, and Clinical Impact of Formula-Based Mass-to-Molar Conversion.The journal of applied laboratory medicine · 2026
- Intra-individual variability in lipoprotein(a): insights from 922,304 measurementsEuropean Heart Journal · 2026
- B-300 Validation of the Lp(a) Ultra Assay on Beckman Coulter AU680: Overcoming Isoform Heterogeneity Challenges in Lipoprotein(a) MeasurementClinical Chemistry · 2026
- Lipoprotein(a) measurement: challenges and opportunities.Clinical chemistry and laboratory medicine · 2026