A-017 Identifying and Resolving Outliers in Lipoprotein(a) Immunoassay Testing

Alexandria Jackson, Jonathan Genzen, Joseph Rudolf, Sara Wyness

Key points

From the paper's conclusion
  • 0.2% of these pairs (n=49) were classified as outliers.

Abstract

Background Lipoprotein(a) [Lp(a)] is an LDL-like particle which expresses ApoB-100 and Apo(a). Elevated Lp(a) is associated with increased risk of atherosclerotic cardiovascular disease (ASCVD) and stroke. Lp(a) is commonly measured on using latex-enhanced immunoturbidimetric methods. As immunoturbidimetry depends on light scattering and antigen-antibody interaction, it can be susceptible to optical, reagent, and matrix interferences. To resolve sporadic, discrepant results (i.e., outliers) occasionally observed in testing for Lp(a), the following studies were performed. Methods Using an IRB-exempt protocol, a dataset sampling period in which duplicate Lp(a) testing was performed on clinical specimens [Roche Tina-quant Lp(a) Gen.2 assay on c503 (Roche) instruments; mg/dL] was retrieved. Additional Lp(a) experiments included evaluation of the Roche assay on c503 instruments and the Randox and Diazyme Lp(a) assays on both c503 and AU5800 (Beckman Coulter) platforms. These subsequent studies used previously tested, de-identified patient specimens, including a subset that previously demonstrated lack of reproducibility. Outliers were defined based on the average of the original and repeat results: absolute differences of greater than or equal to ± 6 mg/dL (± 15 nmol/L) for averages < 30 mg/dL (< 75 nmol/L), or percent differences of = 20% for averages = 30 mg/dL (= 75 nmol/L). Percent difference was calculated as |[(Replicate 1 - Replicate 2) / average of (Replicate 1 and Replicate 2)]| × 100. Replicates were run on the same instrument or instruments of the same model. Replicate pairs where one or both results were outside the analytical measuring range (AMR) were excluded. Results Of 26,637 specimens for which duplicate results were extracted, 20,832 had both original and repeat results within the AMR. 0.2% of these pairs (n=49) were classified as outliers. Instrument service and review of pre-analytical processes (e.g., centrifugation, mixing, random-access versus batching testing) did not eliminate sporadic discrepancies. Clinically, discrepancies were resolved by additional replicate testing. See Table 1 for additional Lp(a) experiments that were conducted and their findings. Assay optimization using Diazyme reagent was then conducted to ensure thorough onboard specimen pipetting, sample/reaction mixing, and washing, yielding decreases in the number of outliers. Conclusion Immunoturbidimetric Lp(a) assays can be subject to outliers that are resolvable with optimization of reagent/sample mixing and washes. As Lp(a) measurement is typically ordered once for ASCVD risk stratification, assay imprecision may negatively impact clinicians’ understanding of patients’ Lp(a) concentrations. With Lp(a)-altering therapies currently under development, ensuring an accurate measurement is essential when considering therapeutic interventions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers