Vial puncture mechanics with a mechanical assist device during anticancer drug compounding with a closed-system drug transfer device.
Yoshiaki Kikuchi, Shun Ishikawa, Toshihisa Nakashima, Daisuke Watabe, Tomohisa Asano, Yoshimasa Saito, Hironobu Hashimoto
Department of Pharmacy, National Cancer Center Japan Hospital, Tokyo, Japan.
Key points
From the paper's conclusion- The assist device for CSTDs stabilized puncture trajectory through the rubber stopper, yielding a more perpendicular and controlled intrastopper pathway than manual puncture. The practical significance of this trajectory stabilization for rubber stopper push-in and occupational exposure remains to be determined.
Abstract
PURPOSE: Off-center and nonperpendicular vial puncture has been proposed as a potential contributor to rubber stopper push-in during needle insertion into closed-system drug transfer devices (CSTDs). However, quantitative evaluation of the puncture process remains limited. This study aimed to compare puncture mechanics during vial access between manual puncture and use of a mechanical assist device and to evaluate differences in the device effect across vial types. METHODS: Anticancer drug vials were compounded using NEOSHIELD with or without NS Spike Assist. Vials were randomly sampled (maximum four per compounder), targeting 70 vials per vial type in each study group. Puncture distance from center (D1), puncture angle (A1), and radial deviation within the rubber stopper (drad) were measured microscopically. RESULTS: A total of 980 vials were analyzed. The assist device shifted A1 toward perpendicular across most vial types and mitigated angular deflection associated with off-center entry (larger D1), with A1 clustering near 90° across a wide range of D1 values. Changes in D1 varied by vial type. drad was predominantly negative with manual puncture but shifted toward neutral or positive with the assist device in several vial types; for doxorubicin 50 mg vials, drad increased from -0.046 to 0.031 mm (ΔHL 0.079 mm, 95% CI 0.021-0.146; P = 0.008). CONCLUSIONS: The assist device for CSTDs stabilized puncture trajectory through the rubber stopper, yielding a more perpendicular and controlled intrastopper pathway than manual puncture. The practical significance of this trajectory stabilization for rubber stopper push-in and occupational exposure remains to be determined.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Study on the residual contents of empty drug vials from intelligent dispensing robots and manual compounding of cytotoxic drugs in PIVAS.Journal of oncology pharmacy practice : official publication of the International Society of Oncology Pharmacy Practitioners · 2026
- Swinging vs Flicking: Assessing Safer Removal Techniques for Air Bubbles in Intravitreal Injections.Journal of vitreoretinal diseases · 2026
- Ready-to-Use Versus Traditional IV Piggyback Preparation: A Combined Process-Error, Workforce-Time-Waste, and Cost-Effectiveness Analysis.Pharmacy (Basel, Switzerland) · 2026
- Preparation of B. Braun Easypump® II elastomeric devices with Chemfort™ closed system transfer devices (CSTDs) in uncontrolled environments.JAC-antimicrobial resistance · 2026
- Protein aggregation in self-lubricating piston pumps: The impact of sustained shear stress.International journal of pharmaceutics · 2026
- Exploring the occupational protection potential of an intravenous compounding robot across the cytotoxic drug workflow.Frontiers in public health · 2026