Preparation of B. Braun Easypump® II elastomeric devices with Chemfort™ closed system transfer devices (CSTDs) in uncontrolled environments.
Louisa Knowles, Abi Jenkins
Pharmacy Department, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK.
Key points
From the paper's conclusion- Preparation using Chemfort™ CSTDs was associated with reduced microbial contamination and shorter preparation times compared with needle-based methods. CSTDs may provide a useful risk-reduction strategy for nurse-prepared elastomeric infusions in OPAT services, although further evaluation in larger studies is required.
Abstract
BACKGROUND: Outpatient parenteral antimicrobial therapy (OPAT) enables patients to receive intravenous antimicrobials outside hospital settings. Elastomeric infusion devices facilitate continuous antibiotic delivery; however, limited stability data and constrained aseptic unit capacity have led some services to adopt nurse-led 'fresh-fill' preparation in uncontrolled environments, which may increase microbial contamination risk. OBJECTIVES: To evaluate whether preparation of B. Braun Easypump® II elastomeric devices using Chemfort™ closed system transfer devices (CSTDs) in uncontrolled environments reduces microbial contamination compared with preparation using needles and syringes. METHODS: Forty-six elastomeric devices were prepared with tryptic soy broth by eight operators across three uncontrolled rooms using either Chemfort™ CSTDs (n = 23) or needles and syringes (n = 23). Aseptic non-touch technique was followed. Devices were incubated at 32°C for 7 days before microbiological testing. Environmental monitoring included settle plates and finger dabs. Fisher's exact test was used to analyse contamination rates. RESULTS: No contamination occurred in devices prepared using CSTDs (0/23), whereas 5/23 (22%) prepared using needles were contaminated. This difference was statistically significant (P = 0.0491). Preparation times were shorter and more consistent using CSTDs (median 26 min; range 20-40) compared with needles and syringes (median 34 min; range 22-64). CONCLUSIONS: Preparation using Chemfort™ CSTDs was associated with reduced microbial contamination and shorter preparation times compared with needle-based methods. CSTDs may provide a useful risk-reduction strategy for nurse-prepared elastomeric infusions in OPAT services, although further evaluation in larger studies is required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of Closed-System Transfer Devices in Reducing Surface Contamination With Antineoplastic Drugs in Two Patient Administration Areas.Clinical journal of oncology nursing · 2026
- Head-to-head containment performance testing of four commercially available closed-system transfer devices using the 2016 Draft NIOSH test protocol and 2.5% w/v 2-phenoxyethanol as challenge agentPLoS ONE · 2026
- Comparative Surface Microbial Contamination of a Cardboard Spacer MDI PLUS® and Conventional Plastic Inhalation Devices under Simulated Repeated Use.The Journal of asthma : official journal of the Association for the Care of Asthma · 2026
- Sterility of Tranexamic Acid Syringes Prefilled by Tactical Combat Medics in a Garrison Setting.Journal of special operations medicine : a peer reviewed journal for SOF medical professionals · 2026
- Comparative Evaluation of Microbial Transfer in Intermittent Catheters: An in vitro Study of Protective Features Against Uropathogens.Journal of wound, ostomy, and continence nursing : official publication of The Wound, Ostomy and Continence Nurses Society · 2026
- Ready-to-Use Versus Traditional IV Piggyback Preparation: A Combined Process-Error, Workforce-Time-Waste, and Cost-Effectiveness Analysis.Pharmacy (Basel, Switzerland) · 2026