69 Identifying the Analytical Interferent of Enzymatic Creatinine Measurements in a Case of Acetaminophen Overdose
Chelsea Swartchick, Brienne Berent, Christopher Rosenmark, Darci Block, Nikola Baumann, Brooke Katzman
Key points
From the paper's conclusion- These findings demonstrate for the first time that NAPQI is a clinically significant interferent in enzymatic creatinine assays and cross-reacts with acetaminophen quantification.
Abstract
The goals of this study were to: i) investigate acetaminophen–related interference with enzymatic creatinine assays in serum specimens from a patient presenting to the Emergency Department with acetaminophen overdose and ii) identify the interfering agent through in vitro spiking and recovery studies. Residual, waste serum specimens that were obtained for physician-ordered testing from a patient with a peak acetaminophen concentration of 588 µg/mL (Roche c502, Roche Diagnostics, Inc., therapeutic interval: 10–30 mcg/mL) were used to measure creatinine across three methods enzymatic Trinder assay (CREAEnzR, Roche c701, Roche Diagnostics, Inc.), kinetic Jaffe reaction (CREAJ, Roche c701, Roche Diagnostics, Inc.), and a dry-slide enzymatic platform (CREAEnzV, Vitros 350, QuidelOrtho Corp.) known to reduce endogenous and exogenous interferences. For residual patient samples (n = 5), CREAJserved as the reference method, against which enzymatic assays (CREAEnzRand CREAEnzV) were evaluated. Enzymatic measurements were considered clinically significant when differences exceeded ± 0.30 mg/dL or 10% compared to CREAJ. A 1.00 mg/dL creatinine serum pool (adult reference interval: 0.59–1.04 mg/dL) was spiked (<10% v/v saline solution) with clinically relevant concentrations of acetaminophen (25–600 µg/mL) or overdose antidote, N-acetyl cysteine (NAC) (50–200 µg/mL). Creatinine was measured with all methods to evaluate recovery. Another serum creatinine pool (0.94 mg/dL) was spiked with reactive metabolite, N-acetyl-p-benzoquinone imine (NAPQI) (20–123 µg/mL, 6% v/v DMSO solution) and tested using CREAEnzRand CREAJ. For spiking experiments, recovery was determined relative to blank matrix-matched controls. Analysis of specimens from a patient with acetaminophen overdose revealed method-specific differences in creatinine measurements. Creatinine was reported as 0.87 ± 0.06 mg/dL in the patient sample using CREAJ, matching pre-admission historical creatinine measurements. In the same samples, creatinine measured 0.36 ± 0.19 mg/dL and 0.43 ± 0.16 mg/dL using the CREAEnzRand CREAEnzVmethods, respectively. Acetaminophen and NAC spiking experiments did not affect creatinine recovery across all methods (97-101% recovery). NAPQI produced a concentration-dependent reduction in creatinine values using CREAEnzR, with recoveries of 49 ± 3% (n = 3) at 75 mcg/mL NAPQI, while creatinine recovery was 103 ± 2% (n = 3) using CREAJ.A serum pool without acetaminophen and spiked with NAPQI resulted in a false elevation of acetaminophen (13 ± 0 mcg/mL) at 75 mcg/mL NAPQI. These findings demonstrate for the first time that NAPQI is a clinically significant interferent in enzymatic creatinine assays and cross-reacts with acetaminophen quantification. Laboratory professionals should be aware of NAPQI-mediated bias in specimens from patients with acetaminophen toxicity, consider alternative measurements of renal function such as Jaffe creatinine methods or cystatin C. Recognizing analytical interference is essential for accurate laboratory interpretation and optimal patient care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Association of serum acetaminophen levels with hepatorenal toxicity: A prospective observational study in Tripura, India.Indian journal of pharmacology · 2026
- A-411 Evaluation of Point-of-Care Creatinine Analyzers: eGFR Categorization Accuracy for Safe Contrast AdministrationClinical Chemistry · 2026
- 132 Improving Phenytoin Therapeutic Drug Monitoring Accuracy Using Six Sigma MetricsAmerican Journal of Clinical Pathology · 2026
- NAPQI adducts in patients with selective hypersensitivity to acetaminophenFrontiers in Pharmacology · 2026
- A-412 To hydrate or not to hydrate? A comparison of the Siemens epoc, Abbott i-STAT, and Beckman Coulter DxC700AU IDMS traceable creatinine assays’ inter-platform variabilityClinical Chemistry · 2026