382 Effect of Transient Thermal Exposure on Photometric Rheology-Based Erythrocyte Sedimentation Rate Results: A Sample Stability Assessment
Megan McCutcheon, Steve Giarrusso, Rachel Whiting
Key points
From the paper's conclusion- Conclusion Conclusions: Transient heat exposure at 95 °F for 4 hours does not produce clinically significant changes in iSED ELITE ESR results, supporting specimen reliability despite short-term heat exposure prior to testing.
Abstract
Introduction/Objective Introduction: Elevated temperatures during specimen transport may affect erythrocyte sedimentation rate (ESR) results, however, the impact of transient heat exposure prior to return to room temperature on ESR accuracy has not been well characterized. Unlike other ESR methodologies, the iSED® family of analyzers has 28 hours of room temperature ESR sample stability (vs. 4 hours afforded by traditional ESR methods), yet maintaining room temperature storage can be a challenge in warmer climates when collection sites are widely dispersed. Methods/Case Report Methods: Residual whole blood specimens (n = 245) were analyzed for ESR at room temperature (18–25 °C) using the iSED ELITE analyzer (ALCOR Scientific, Smithfield, RI), incubated at 95 °F for 4 hours, returned to room temperature, and re-analyzed. Agreement was assessed by intraclass correlation coefficient (ICC) with 95% confidence intervals. Artificial intelligence tools were used to assist with statistical analysis. Results Results: Baseline mean ESR was 39.6 ± 25.4 mm/hr; post-heating mean was 38.9 ± 23.6 mm/hr. The mean difference was −0.8 mm/hr (95% CI: −1.6 to 0.1). ICC for absolute agreement was 0.960 (95% CI: 0.95–0.97), indicating excellent concordance. Conclusion Conclusions: Transient heat exposure at 95 °F for 4 hours does not produce clinically significant changes in iSED ELITE ESR results, supporting specimen reliability despite short-term heat exposure prior to testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Measuring Erythrocyte Sedimentation Rate Using Photometric Rheology Technology Demonstrates Superior Sample Stability as Compared to the Westergren Method.Diagnostics (Basel, Switzerland) · 2026
- Performance Evaluation of the Alifax Test1 2.0 for Erythrocyte Sedimentation Rate Testing.Clinical laboratory · 2026
- Effects of storage temperature and delayed analysis on routine biochemical and haematological test resultsScientific Reports · 2026
- Too slow Erythrocyte Sedimentation Rate: Deeper biophysical understanding, novel accurate parameters and new medical applicationsmedRxiv · 2026
- Quality Indicator-Driven Six Sigma Scale Assessment of Extra-Analytical Phases in Clinical Chemistry Laboratory-Insights from a Tertiary Care Hospital.Annals of clinical biochemistry · 2026
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026