Association between FMEA occurrence rankings and linear accelerator maintenance episode counts.
Min Ma, Ke Zhang, Hongkai Wang, Jianrong Dai
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, People's Republic of China.
Key points
From the paper's conclusion- We compared occurrence ranks from 2017-2021 with episode counts from 2018-2021 across 36 accelerator and QC item units using Spearman correlation, a QC item cluster bootstrap, block permutation, and Kendall tau-b.
Abstract
Objective.Failure mode and effects analysis (FMEA) guides quality control (QC) priorities, but occurrence rankings based on tolerance exceedances may not represent corrective maintenance. We assessed whether higher historical QC occurrence ranks corresponded to more corrective maintenance episodes in the same QC domain.Methods.We analyzed 643 maintenance records and 30 523 numerical QC observations from three accelerators. For each accelerator, consecutive corrective maintenance entries with the same documented mechanism and gaps of no more than three calendar days were grouped into an episode under the 72 h rule. A fixed crosswalk assigned eligible episodes to 12 historical FMEA items through direct or indirect numerical mappings. We compared occurrence ranks from 2017-2021 with episode counts from 2018-2021 across 36 accelerator and QC item units using Spearman correlation, a QC item cluster bootstrap, block permutation, and Kendall tau-b. Sensitivity analyses examined the unbinned occurrence probabilities, episode definitions, and exclusion of the multileaf collimator (MLC) domain.Results.The primary maintenance period contained 393 eligible episodes. Of 224 episodes with numerical representation, 213 (95.1%) had only indirect or surrogate mappings; 209 entered the 12 item crosswalk. The QC worksheets contained 41 confirmed tolerance exceedances in 10 of 36 units. Laser localization accounted for 18 exceedances but no mapped maintenance episode. MLC leaf position accounted for two exceedances and 137 mapped episodes, including 74 leaf drive motor failures. Spearman correlation was 0.19 (95% QC item cluster resampling interval, -0.36 to 0.65); Kendall tau-bwas 0.17 (95% interval, -0.33 to 0.57). Numerical coverage was absent for 169 episodes (43.0%).Conclusion.Historical occurrence ranks showed a weak, imprecisely estimated association with maintenance episode counts. Differences between QC domains indicate a need to check how QC measures relate to documented maintenance mechanisms before using these scores to represent maintenance activity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detecting the Failures Nobody Sees: Applying Pareto Analysis and Failure Mode and Effects Analysis to Equipment Reliability in a Healthcare Simulation Institute.Cureus · 2026
- Application of an updated severity-based FMEA approach to Radiation Therapy QA: A multi-institutional exercise.Journal of applied clinical medical physics · 2026
- A-219 Risk-Based Ranking of Microscopists in Manual Leukocyte Differentials: A Personalized FMEA Linking Precision, Trueness, and Accuracy Indices to Rumke Interval NonconformitiesClinical Chemistry · 2026
- Penerapan FMEA untuk peningkatan keandalan perawatan mesin las MIG di Perusahaan XYZJENIUS Jurnal Terapan Teknik Industri · 2026
- Evidence Before Optimization: Maintenance Prioritization by Evidence Sufficiency in Compressed-Air SystemsApplied Sciences · 2026
- Automated linac downtime tracking across 4 radiation oncology clinics within a single hospital system utilizing statistical process control.Journal of applied clinical medical physics · 2026