Automated linac downtime tracking across 4 radiation oncology clinics within a single hospital system utilizing statistical process control.
Caroline E Serafini, Matthew Thomas, Xiaofeng Yang, Pretesh Patel, Justin Roper, Maxine Washington, Anees Dhabaan, Hania Al-Hallaq
Department of Radiation Oncology, Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.
Key points
From the paper's conclusion- SPC applied to routinely collected operational data provides a scalable framework for automated linac downtime detection. Integration of automated monitoring with structured reporting may improve downtime surveillance and inform proactive operational planning in radiation oncology.
Abstract
BACKGROUND: Linear accelerator (linac) downtime disrupts clinical workflow and may delay patient treatment; however, downtime events are frequently undocumented or recorded in informal reporting systems. Data-driven approaches using routinely collected operational metrics may improve downtime surveillance and provide insight into system-level adaptation. PURPOSE: This study aims to evaluate if statistical process control (SPC) methods can be applied to routine treatment and scheduling data from an oncology information system (OIS) to identify linac downtime events and characterize operational response across multiple radiation oncology clinics within a single hospital system. METHODS: Daily machine-level operational data from 10 linacs across 4 clinics were analyzed over 29 months (January 2024-May 2026). Machine metrics included daily patient volume, cancellations, and operating hours. For each machine, rolling 45-day baseline means and standard deviations were computed, and SPC rules (± 3σ control limits and > 10 cancellations / day) were applied to detect downtime days. SPC-flagged events were compared with email-based downtime event logs to calculate sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). Downtime events were categorized by cause, and intra-hospital patient redistribution was assessed by identifying concurrent > 4σ volume increases on alternate machines. RESULTS: Over the 29-month study period, 138 downtime events were documented in downtime event logs. SPC identified 23 additional events exhibiting operational signatures consistent with downtime. Using reported events as the reference standard, SPC achieved sensitivity 0.70, specificity 0.99, PPV 0.79, and NPV 0.99. Downtime frequently resulted in compensatory increases in patient volume on alternate machines, with 41 transfer pairs identified across three hospitals. When the reported downtime cause was specified, machine mechanical causes accounted for the largest proportion (48%) of downtime events. CONCLUSIONS: SPC applied to routinely collected operational data provides a scalable framework for automated linac downtime detection. Integration of automated monitoring with structured reporting may improve downtime surveillance and inform proactive operational planning in radiation oncology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Linac performance benchmarking: Feasibility of a web-based tool for comparison of local machine performance check (MPC) results with the global community.Journal of applied clinical medical physics · 2026
- Unplanned radiotherapy equipment downtime: Impact on clinical and organizational operations.Journal of medical imaging and radiation sciences · 2026
- Data-driven Assessment of Magnetic Resonance Imaging (MRI) Utilization and Aggregate Electrical Demand in U.S. Healthcare Facilities.Academic radiology · 2026
- Analysing workflows in radiation oncology: A process mining approach to identify treatment process challenges.Tumori · 2026
- Daytime Idle Energy Reduction in Radiotherapy Simulation Computed Tomography: A Prospective Workflow-Based Evaluation.Journal of medical radiation sciences · 2026
- B-019 Operational and Economic Impact of Turnaround Time Target Exceedance in a High-Volume Automated Clinical LaboratoryClinical Chemistry · 2026