When Bioinspiration Doesn't Help: Mechanical Behavior of Bio-inspired Curved Needles During Curvilinear Insertion.
Doyoung Kim, Parsaoran Hutapea
Department of Mechanical Engineering College of Engingeering, Temple University, 1947 N 12th Street, Philadelphia, Pennsylvania, 19122, United States.
Key points
From the paper's conclusion- These findings suggest that, for curved needles following predefined trajectories under the tested conditions, the measured forces did not change significantly with the addition of barbed surface features under curvilinear insertion.
Abstract
This study investigates the effectiveness of bio-inspired barbed geometries applied to curved needles for percutaneous procedures. The geometries of insect stingers, such as honeybees and mosquitoes, have been applied to straight needles to reduce insertion forces, tissue deformation, and tissue damage during linear insertion. However, whether these benefits extend to curved needles undergoing curvilinear base manipulation remains unclear. In this study, mosquito-inspired barb geometry was adapted onto bevel-tip curved needles with bent angles of 7.5°, 15°, 22.5°, and 30°. The effect on transverse and axial insertion forces, bending moments, and needle-tip trajectories was evaluated. The regular and bio-inspired curved needles were fabricated using high-resolution 3D printing and tested in tissue-mimicking gel under curvilinear insertion. Force and moment data were recorded at the needle base, and the needle tip paths were quantified using radius of curvature analysis from video tracking. No statistically significant reductions were found in insertion forces or bending moments under curvilinear insertion. These findings suggest that, for curved needles following predefined trajectories under the tested conditions, the measured forces did not change significantly with the addition of barbed surface features under curvilinear insertion. This suggests that the tested surface modification does not substantially alter the overall mechanical response under these conditions. This work highlights that friction-reducing surface features effective in linear insertion may not directly translate to curvilinear insertion, and that optimization strategies for curved needle applications may benefit from prioritizing geometric and trajectory-based design considerations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-bevel steerable needles: A mechanics-driven framework for simultaneous optimization of deflection, insertion force, and penetration safety.Journal of the mechanical behavior of biomedical materials · 2026
- Friction and interfacial mechanics of surgical suturing: needle–tissue penetration, suture drag and friction-control strategiesFrontiers in Mechanical Engineering · 2026
- Planar Bending of Flexible Instrument Shafts: Modeling and Experimental ValidationCurrent Directions in Biomedical Engineering · 2026
- Design Improvement of a 5-Bar Linkage-Based Needle Insertion Robot for Pain Intervention ProceduresMachines · 2026
- Biomechanical study of probe-tissue interactions during deep brain stimulation: Force measurement and channel deformation in agar gel phantoms.Journal of the mechanical behavior of biomedical materials · 2026
- Principles of use of tensile J-curve materials in antagonistic arrangementsProceedings of the National Academy of Sciences · 2026