Medical engineering & physics

Force-Torque Classification of Reference-Profile-Defined Drilling States During Robot-Assisted Dental Implant Osteotomy.

Jiaqiang Li, Yangdong Wu, Yao Wang, Min Shao

Key Laboratory of Advanced Manufacturing Technology, Guizhou University, Huaxi campus, No.515,Jiaxiunanlu Road, Guiyang, Guizhou, 550025, China. · Key Laboratory of Advanced Manufacturing Technology of the Ministry of Education, Guizhou University, No.515 Jiaxiu South Road, Guiyang, Guizhou Province, P.R.China, 550025, Guiyang, Guizhou, 550025, China.

Key points

From the paper's conclusion
  • To evaluate whether force-torque features can classify three predefined, reference-profile-defined operational drilling states during robot-assisted implant osteotomy under specimen-level separation.
Materials and Methods: Thirty porcine hemimandibular specimens underwent 180 robot-controlled osteotomies under three drilling conditions.

Abstract

To evaluate whether force-torque features can classify three predefined, reference-profile-defined operational drilling states during robot-assisted implant osteotomy under specimen-level separation. Materials and Methods: Thirty porcine hemimandibular specimens underwent 180 robot-controlled osteotomies under three drilling conditions. Axial force (Fz) and drilling torque (Mz) were sampled at 1000 Hz, segmented into 100-ms windows with a 50-ms step, and represented by 42 time-, frequency-, and wavelet-domain features; 23 features were selected using training-only ANOVA ranking and correlation pruning. Specimens were separated into training, validation, and held-out test sets (18/6/6). Linear support vector machine (SVM), random forest (RF), and XGBoost were evaluated against predefined depth-indexed operational reference profiles; a same-acquisition CBCT rule served as an imaging comparator. Results: The dataset contained 92,675 overlapping windows. On six held-out specimens (36 osteotomies; 18,496 windows), XGBoost achieved 93.0% accuracy (specimen-clustered 95% CI, 92.0-94.0%), 82.9% balanced accuracy, and 85.6% macro-F1. RF and SVM accuracies were 90.5% and 88.8%, respectively. The CBCT rule achieved 92.3% accuracy but lower balanced accuracy (80.1%) and transitional recall (51.1%). XGBoost transitional recall was 59.6%, increasing to 73.0% in a weighted sensitivity configuration at the cost of lower overall accuracy. Accuracy decreased from 94.1% away from class boundaries to 75.6% in boundary-adjacent windows. Conclusion: Force-torque features differentiated the three predefined reference-profile-defined drilling states under specimen-separated evaluation. XGBoost achieved the highest classification performance, with errors concentrated primarily in transitional and boundary-adjacent windows. Keywords: robot-assisted implant osteotomy; force-torque sensing; reference-profile classification; drilling-state classification; XGBoost; specimen-level validation&#xD.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers