Radiation monitoring in the scattered field of a clinical electron FLASH beam with Timepix4
Andrea Garcia-Tejedor, Walter Reiner Geyer, Riccardo Bolzonella, Pierre Carbonez, Jerome Damet, Claude Bailat
Key points
From the paper's conclusion- This work reports a first evaluation of a Timepix4 assembly bonded to a 300 μm p-on-n sensor in the scattered field of a clinical electron beam, inside and outside of the treatment room, at the Mobetron facility (CHUV, Lausanne), using a 9 MeV beam delivered at conventional (∼ 0.5 Gy/s) and high (30–230 Gy/s) dose rates.
Abstract
FLASH radiotherapy is an emerging technique, delivered at ultra-high dose rates, that holds promise for improving the treatment of certain cancers. Resolving the pulse structure of the beam is a major challenge in the primary field, where the extreme particle flux prevents accurate pulse-resolved measurements. Even in the scattered field, few instruments provide the precision required for radiation protection and beam monitoring. Single-particle counting detectors offer a way forward: the Timepix family of hybrid pixel detectors records each particle interaction individually, with its time of arrival and time over threshold. This work reports a first evaluation of a Timepix4 assembly bonded to a 300 μm p-on-n sensor in the scattered field of a clinical electron beam, inside and outside of the treatment room, at the Mobetron facility (CHUV, Lausanne), using a 9 MeV beam delivered at conventional (∼ 0.5 Gy/s) and high (30–230 Gy/s) dose rates. Individual beam pulses were resolved and counted in both modes, with repetition frequencies consistent with the machine settings. Within high dose rate pulses, progressive data loss from readout saturation distorts the recorded pulse shape, so the intra-pulse profile is not reliable, while pulse counting and timing are unaffected. Outside the treatment room, where passive dosimeters record only background, individual pulses were still resolved in high-dose-rate mode. These results establish Timepix4 as a viable pulse-resolved detector for radiation protection and out-of-field monitoring in FLASH RT, reliably resolving pulse timing and counts even at dose levels below what passive dosimeters can register.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulse-resolved spatial dosimetry of ultra-high dose rate electron FLASH beams using optical imaging.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- A scintillator-based approach to spatial and temporal dosimetry in UHDR electron irradiationJournal of Physics: Conference Series · 2026
- Study of three-dimensional dose distribution measurement of ultra-high dose rate electron beam.Physics in medicine and biology · 2026
- Ultra-high-dose-rate pulsed irradiation and the conditional FLASH effect: dosimetry, multiscale biology, and clinical translation.Frontiers in oncology · 2026
- Dosimetric Characterization and Workflow Optimization of the FLASH-SARRP for Reliable Preclinical Radiobiological StudiesbioRxiv · 2026
- Charge collection efficiency in air-vented plane-parallel ionisation chambers at ultra-high dose rates: a self-consistent Garfield++ Monte Carlo model including space-charge effects and ion recombinationPhysics in Medicine and Biology · 2026