Journal of Physics: Conference Series

Development of a 3D Slicer Extension for MR-Based Gel Dosimetry Analysis

G S Lee, K M Alexander, JD Jutras

Key points

From the paper's conclusion
  • In this work, an MR-based 3D Slicer gel dosimetry extension was developed with MR-specific analysis tools for ΔR1 or ΔR2 mapping and MR-based dose calibration.

Abstract

With the increasing adoption of MR-guided radiation therapy systems, there is renewed interest in MR-based gel dosimetry readout and a need for dedicated analysis tools. In this work, an MR-based 3D Slicer gel dosimetry extension was developed with MR-specific analysis tools for ΔR1 or ΔR2 mapping and MR-based dose calibration. The workflow was demonstrated using a Fricke gel irradiated with a liver SBRT plan. 3D gamma analysis of the MR readout yielded a pass rate of 96.85% using 2%/2 mm criteria when compared with the planned dose distribution, while good agreement was also observed with optical CT measurements. The extension provides an efficient method for MR-based gel dosimetry analysis.

Read the paper