Value transmission reliability of the standard platinum resistance thermometer based on the metrological traceability system and its application in infrared temperature ear thermometer calibration
Liwei Zhang, Chunlin Cheng, Wenwen Wang, Zhi Chen, Shiwei Xu
Key points
From the paper's conclusion- The research results indicate that by strictly following national verification regulations (such as JJG 160-2007 Verification Regulation of Standard Platinum Resistance Thermometer and JJG 1164-2019 Verification Regulation of Infrared Ear Thermometers ), implementing standardized periodic traceability, annealing treatment, and stability monitoring, the uncertainty contribution of platinum resistan
Abstract
In modern metrology systems, accurate transmission of temperature values is the cornerstone of ensuring comparability and reliability of industrial production, scientific research, and public health monitoring data. The platinum resistance thermometer (PRT) serves as the core carrier of the International Temperature Scale (ITS-90), and its metrological performance directly determines the effectiveness of the temperature value traceability chain. This article starts from the perspective of the national metrological traceability system and systematically studies the key factors that affect the reliability of value transmission of the standard platinum resistance thermometer, including long-term stability, stress-induced drift, oxidation effect, and thermal installation error. The relevant effects are quantitatively analyzed through experimental data and metrological specifications. Meanwhile, this article takes the calibration of infrared ear thermometers, a typical medical measuring instrument, as an example to construct a complete traceability chain model using a second-class standard platinum resistance thermometer as a medium, and evaluates in detail the various traceability-related components in its measurement uncertainty. The research results indicate that by strictly following national verification regulations (such as JJG 160-2007 Verification Regulation of Standard Platinum Resistance Thermometer and JJG 1164-2019 Verification Regulation of Infrared Ear Thermometers ), implementing standardized periodic traceability, annealing treatment, and stability monitoring, the uncertainty contribution of platinum resistance thermometer as a transmission standard can be minimized, thereby ensuring the accuracy and uniformity of the measurement values of terminal temperature measuring equipment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Minimising the uncertainty in T -T90 above the silver pointMeasurement Science and Technology · 2026
- Investigation of the active laser pyrometer with reflection for platinum dynamic surface true temperature measurement.Optics letters · 2026
- Reliability and Agreement of Infrared Thermography and a Digital Laser Thermometer for Non-Invasive Body Temperature Assessment in SowsAnimals · 2026
- Assessment of the calibration capability of KRISS for noble-metal thermocouplesMetrologia · 2026
- Metrological Validation of Low-Cost DS18B20 Digital Temperature Sensors Using the TH-001 Procedure: Calibration Models, Uncertainty, and ReproducibilityMetrology · 2026
- Conversion of Reported Temperatures from Historical Scales to the ITS-90International Journal of Thermophysics · 2026