Assessment of the calibration capability of KRISS for noble-metal thermocouples
Yong-Gyoo Kim, Young Hee Lee, Sunghun Kim, Suyong Kwon
Key points
From the paper's conclusion- To establish the reliability of this practical approach, a comprehensive assessment of KRISS’s thermocouple calibration capability was undertaken.
Three Type R reference thermocouples were prepared, characterized for thermoelectric inhomogeneity, and calibrated at the ITS-90 fixed points (Sn, Zn, Al, Ag, Cu), and Pd melting point realized by the KRISS cell together with selected metal–carbon e
Abstract
The Korea Research Institute of Standards and Science (KRISS) maintains the national temperature standards and provides high-accuracy calibration services exclusively for noble-metal thermocouples (Types R, S and B). Most industrial users prefer comparison calibration to primary fixed-point calibration because the latter is time-consuming and costly. Comparison calibrations, however, carry larger uncertainties, and values above 1100 °C must be obtained by extrapolation, which further increases the uncertainty. To establish the reliability of this practical approach, a comprehensive assessment of KRISS’s thermocouple calibration capability was undertaken. Three Type R reference thermocouples were prepared, characterized for thermoelectric inhomogeneity, and calibrated at the ITS-90 fixed points (Sn, Zn, Al, Ag, Cu), and Pd melting point realized by the KRISS cell together with selected metal–carbon eutectic points (Fe–C, Co–C and Pd–C) up to 1560 °C. The same thermocouples were then mutually intercompared in a fully automated comparison system up to 1100 °C, after which the deviation functions were extrapolated to 1600 °C. Temperature differences between the extrapolated comparison results and the independent fixed-point calibrations remained within ±0.3 °C across the entire range—substantially smaller than the expanded uncertainties assigned to routine comparison calibrations (0.8 °C from the ice point to 1100 °C and 1.6 °C from 1100 °C to 1600 °C). These results demonstrate that fixed-point-calibrated reference thermocouples can be used for routine comparison calibrations limited to 1100 °C with subsequent extrapolation to 1600 °C, while remaining fully consistent with the primary temperature scale. The combination of primary fixed-point realization, mutual inter-comparison, and validated extrapolation therefore provides a practical and fully traceable foundation for KRISS’s thermocouple calibration services.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reversible and irreversible thermoelectric inhomogeneities in type S thermocouplesMeasurement Science and Technology · 2026
- Realizations of the triple points of carbon dioxide and sulfur hexafluoride-ITS-90 temperature determinations and interpolations.Metrologia · 2026
- Minimising the uncertainty in T -T90 above the silver pointMeasurement Science and Technology · 2026
- Conversion of Reported Temperatures from Historical Scales to the ITS-90International Journal of Thermophysics · 2026
- Value transmission reliability of the standard platinum resistance thermometer based on the metrological traceability system and its application in infrared temperature ear thermometer calibrationJournal of Physics: Conference Series · 2026
- Study on the long-term stability and consistency of realizations of the triple point of carbon dioxideMetrologia · 2026