Journal of Physics: Conference Series

An investigation into the application of dual venturi flow meters in high-temperature medium environments

Xingdi Cao, Jie Chen, Yue Liu, Lin Zhang, Yongxin Gao, Lanlan Wang, Renyan Zhang, Qinglei He,

Key points

From the paper's conclusion
  • These results confirm the scheme’s effectiveness in mitigating thermal degradation and measurement errors, providing a reliable technical reference for flow monitoring in extreme industrial applications.

Abstract

This study addresses flow measurement challenges in ultra-high-temperature flue gas environments by proposing thermodynamic compensation for dual Venturi flow meters and rigorous instrument room thermal control. A thermal balance model was established to ensure differential pressure transmitters operate within certified limits, leading to their relocation to a controlled environment below 50 °C. Validation under 260 °C operating conditions demonstrated that the optimized system achieves a measurement uncertainty of ±2.00%, satisfying engineering accuracy requirements. These results confirm the scheme’s effectiveness in mitigating thermal degradation and measurement errors, providing a reliable technical reference for flow monitoring in extreme industrial applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

An investigation into the application of dual venturi flow meters in high-temperature medium environments — Science Skim