Measurement Science and Technology

Calibration of Critical flow Venturi Nozzles in the transitional and turbulent boundary layer regime using hydrogen gas and air

Edvardas Venslovas, Maxime Mussard, Hamidou Soumaré, Rémy Maury

Key points

From the paper's conclusion
  • The results show that the discharge coefficients of the CFVNs for hydrogen and air are within the expanded uncertainty of ±0.3 % (k=2) of the laboratory.

Abstract

Critical flow Venturi Nozzles (CFVNs) are expected to play a key role for metrology of gases in the future due to their affordability, reliability and flexibility. While their performance with several inert gases is well documented, there is a lack of experimental data when used with hydrogen. The objective of this work is to estimate the relevance of calibrating CFVNs with air as a surrogate gas for hydrogen, especially at transitional and turbulent flow regimes. Two nozzles with diameters 1 mm and 2 mm are calibrated with both hydrogen gas and air using a primary standard based on a pressure-Volume-Temperature-time (pVTt) system. The results show that the discharge coefficients of the CFVNs for hydrogen and air are within the expanded uncertainty of ±0.3 % (k=2) of the laboratory.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers