Sensitivity of silicon-to-water dose conversion and Bragg-peak metrics to stopping-power datasets in proton dosimetry.
Flávio Matias da Silva, Julian Shorto, Pablo de Vera, Rafael Garcia-Molina, Isabel Abril, T F Silva, Júlio Pereira, Helio Yoriyaz
Centro de Engenharia Nuclear, Instituto de Pesquisas Energéticas e Nucleares, Avenida Professor Lineu Prestes, 2242, São Paulo, SP, 05508-000, Brazil. · Instituto de Pesquisas Energeticas e Nucleares, Avenida Professor Lineu Prestes, 2242, Sao Paulo, SP, 05508-000, Brazil.
Key points
From the paper's conclusion- This study
quantifies how water-to-silicon stopping-power ratios, converted dose-to-water distributions,
and Bragg-peak metrics vary with the stopping-power dataset for a 67.5 MeV pristine proton
beam in water.
Abstract
OBJECTIVE: Accurate proton dosimetry requires internally consistent stopping-power data for detector-to-water dose conversion and Monte Carlo radiation transport simulations. This study quantifies how water-to-silicon stopping-power ratios, converted dose-to-water distributions, and Bragg-peak metrics vary with the stopping-power dataset for a 67.5 MeV pristine proton beam in water. APPROACH: PHITS simulations were performed using dataset-specific stopping powers from SRIM-2013, PSTAR/NIST, ATIMA, and TDDFT-Penn for proton energies up to 10 MeV, combined with a common SBETHE-based high-energy extension above 10 MeV. Normalized percentage depth-ionization data from the PTW silicon diode in the published benchmark of Faddegon et al. (2015) were converted into relative percentage depth- dose distributions in water using depth-dependent water-to-silicon stopping-power ratios. Direct dose-to-water simulations were performed to evaluate Bragg-peak and distal-range metrics. MAIN RESULTS: The water-to-silicon stopping-power ratio exhibited a pronounced depth dependence, with relative entrance-to-distal variations ranging from 17.40% to 22.10% across the datasets. All converted relative PDD curves reached their maximum at 36.86 mm, with no dataset-dependent shift. This position agrees with the experimental Bragg-peak depth of (36.81 ± 0.15) mm, defined as the midpoint between adjacent PDI maxima at 36.76 and 36.86 mm. The converted curves were approximately 3% below the reference curve throughout the entrance and plateau regions, while dataset-dependent deviations of up to 2.1% were observed near the Bragg peak. Direct PHITS calculations predicted Bragg-peak depths ranging from 36.575 to 36.675 mm; only the TDDFT-Penn result lay within the quoted experimental uncertainty. SIGNIFICANCE: For this benchmark, the stopping-power treatment affects the normalization of the converted dose throughout the entrance, plateau, and distal regions more strongly than it affects the Bragg-peak position. Harmonizing stopping-power data between detector-to-water conversion and Monte Carlo transport, or explicitly including dataset dependence in the uncertainty budget, is important for proton-beam calibration, commissioning, quality assurance, and Monte Carlo validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Proposal of a consistent value for the mean excitation energy of liquid water using advanced modeling, detailed simulations and precission proton Bragg curvesarXiv · 2026
- Three-dimensional computational evaluation of water-to-air mass stopping power ratio for therapeutic helium-, carbon-, oxygen-, and neon-ion beams and its application to patient-specific quality assurance.Physics in medicine and biology · 2026
- Angular Distributions of Secondary Neutron and Photon Ambient Dose Equivalent in Proton Therapy: A Systematic PHITS 3.35 Study at 70-250 MeV.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Proton treatment-plan dose verification with a DPM-based Monte Carlo engine.Frontiers in oncology · 2026
- Patient-Specific Stopping Power Calibration of Treatment Planning X-Ray CT Using Integrated Depth-Dose Profiles Acquired by Proton Transmission Imaging - A Feasibility Study.Physics in medicine and biology · 2026
- Noisy probing dose facilitated dose prediction for pencil beam scanning proton therapy: Physics enhances generalizabilityMedical Physics · 2026