In silico feasibility and robustness evaluation of a single-spot proton Bragg peak FLASH platform for small-animal irradiation using a generalized 3D-printed bolus.
Rutger Jan Cornelis de Koster, Jeremy Putra Gunawan, Lara Barazzuol, Marc-Jan van Goethem, Rob P Coppes, Alexander Gerbershagen, Stefan Both
Particle Therapy Research Center (PARTREC), University Medical Centre Groningen, Zernikelaan 25, Groningen, GR, 9747 AA, Netherlands.
Key points
From the paper's conclusion- This work aims to develop and evaluate an irradiation setup for ultra-high dose-rate Bragg peak proton delivery to orthotopic mammary tumours in mice, employable in image-guided research rooms powered by clinical cyclotrons. 
Approach
We built a simulation framework (Python+TOPAS) to assess a single spot, ablative configuration comprising a copper degrader, collimator, and generalised (3D-
Abstract
Objective FLASH radiotherapy has shown potential to reduce normal tissue toxicity while maintaining tumour control, motivating the development of clinical delivery strategies. However, most preclinical proton FLASH studies rely on transmission beams or spread out Bragg peaks, which do not reflect conformal clinical dose distributions. This work aims to develop and evaluate an irradiation setup for ultra-high dose-rate Bragg peak proton delivery to orthotopic mammary tumours in mice, employable in image-guided research rooms powered by clinical cyclotrons. Approach We built a simulation framework (Python+TOPAS) to assess a single spot, ablative configuration comprising a copper degrader, collimator, and generalised (3D-printed) bolus. Performance was investigated on a Phantom Mice Cohort (N=6) using personalised boluses and a Real Mice Cohort (N=10; µCT/PET) using a cohort-derived generalised bolus. Planning goals were D99%,GTV=19.0Gy and HIGTV≤1.30. Physical dose-rate feasibility was evaluated against a nominal 40Gy/s DRmin,skin threshold under an assumed 500nA beam current cap; this metric does not establish a biological FLASH effect. Robustness evaluation included 0.35 and 0.5mm setup errors and ±3% range uncertainties. Main results Simulated plans met target and heterogeneity goals across cohorts, with the 40Gy/s DRmin,skin criterion met under the assumed beam current conditions (Dmean,skin≈18.5-19.3Gy). The modelled dose-rate criterion was met below the 500nA beam current limit in both cohorts. Under the evaluated 0.35mm setup and ±3% range uncertainties, reduction in D99%,GTV remained within 3% for 9/10 real mouse cases. The generalised bolus achieved the simulated dosimetric performance across ten anatomically distinct mice; applicability to independent cohorts remains unestablished. Significance This in silico evaluation supports the simulated dosimetric and physical dose-rate feasibility of conformal Bragg peak proton delivery in small animals using a generalised bolus. The presented framework enables further development in image-guided proton research facilities, forming a strong basis for experimental validation and commissioning toward in situ small animal studies.
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Novel Bragg peak FLASH radiotherapy: treatment of base-of-skull tumors within existing clinical guidelines.Frontiers in oncology · 2026
- Single‐field‐uniform‐dose‐per‐fraction simultaneous dose and dose rate optimization (SFUDPF–SDDRO) method for proton FLASH therapyMedical Physics · 2026
- Dosimetric Characterization and Workflow Optimization of the FLASH-SARRP for Reliable Preclinical Radiobiological StudiesbioRxiv · 2026
- Dynamic Range Modulation via Integrated Universal Range Shifter and Ridge Filter: Enabling Simplified Delivery for FLASH-Enabled IMPT.Physics in medicine and biology · 2026
- Expanding dose and dose-rate capabilities of a synchrotron-based pre-clinical proton FLASH irradiation platform.Medical physics · 2026
- Experimental validation of time-driven proton PBS FLASH delivery on a clinical proton therapy system.Physics and imaging in radiation oncology · 2026