Experimental validation of time-driven proton PBS FLASH delivery on a clinical proton therapy system.

Dawid Krzempek, Jan Gajewski, Hubert Jabłoński, Anna Spaleniak, Wiktor Komenda, Konrad Guguła, Michał Kamiński, Michał Hankiewicz,

Institute of Nuclear Physics, Polish Academy of Sciences, Krakow, PL-31342, Poland.

Key points

From the paper's conclusion
  • Time-driven delivery enables experimental proton PBS operation at UHDR while maintaining beam characteristics suitable for reproducible dosimetry and radiobiological studies, providing a practical and experimentally characterised platform for preclinical proton PBS FLASH research.

Abstract

BACKGROUND AND PURPOSE: Clinical proton pencil beam scanning systems use dose-driven delivery, limiting operation at ultra-high dose rates required for FLASH radiotherapy. Time-driven spot delivery, controlled by irradiation time rather than monitor chamber feedback, may enable FLASH operation. We implemented a non-clinical time-driven FLASH strategy on a clinical proton PBS system and experimentally validated its performance. MATERIALS AND METHODS: FLASH delivery was achieved by increasing beamline transmission through the energy-selection system and replacing dose-driven spot control with time-driven delivery in the Scanning Controller Diagnostic mode. Beam transmission was evaluated at gantry angle 0°. Depth-dose distributions, beam optics, scanning accuracy and uniform field generation were assessed for 228.4 MeV protons at gantry angles of 0°, 180° and 270°. RESULTS: Opening the ESS divergence-limiting slits increased beamline transmission from 8% to 57%, resulting in approximately 273 nA at the isocentre (1.7 × 1012 protons s-1) and an estimated peak on-axis average dose rate of 2.0 × 103 Gy s-1. An 11 × 11 spot field delivered 25 Gy in approximately 0.6 s, corresponding to an average field dose rate of 42 Gy s-1, with up to 90 Gy s-1 measured at higher doses. Proton range and Bragg peak width remained unchanged. Spot position deviations were below 0.5 mm, and uniform 4 × 4 cm2 fields with <3% flatness were achieved. CONCLUSIONS: Time-driven delivery enables experimental proton PBS operation at UHDR while maintaining beam characteristics suitable for reproducible dosimetry and radiobiological studies, providing a practical and experimentally characterised platform for preclinical proton PBS FLASH research.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers