Experimental validation of time-driven proton PBS FLASH delivery on a clinical proton therapy system.
Dawid Krzempek, Jan Gajewski, Hubert Jabłoński, Anna Spaleniak, Wiktor Komenda, Konrad Guguła, Michał Kamiński, Michał Hankiewicz,
Institute of Nuclear Physics, Polish Academy of Sciences, Krakow, PL-31342, Poland.
Key points
From the paper's conclusion- Time-driven delivery enables experimental proton PBS operation at UHDR while maintaining beam characteristics suitable for reproducible dosimetry and radiobiological studies, providing a practical and experimentally characterised platform for preclinical proton PBS FLASH research.
Abstract
BACKGROUND AND PURPOSE: Clinical proton pencil beam scanning systems use dose-driven delivery, limiting operation at ultra-high dose rates required for FLASH radiotherapy. Time-driven spot delivery, controlled by irradiation time rather than monitor chamber feedback, may enable FLASH operation. We implemented a non-clinical time-driven FLASH strategy on a clinical proton PBS system and experimentally validated its performance. MATERIALS AND METHODS: FLASH delivery was achieved by increasing beamline transmission through the energy-selection system and replacing dose-driven spot control with time-driven delivery in the Scanning Controller Diagnostic mode. Beam transmission was evaluated at gantry angle 0°. Depth-dose distributions, beam optics, scanning accuracy and uniform field generation were assessed for 228.4 MeV protons at gantry angles of 0°, 180° and 270°. RESULTS: Opening the ESS divergence-limiting slits increased beamline transmission from 8% to 57%, resulting in approximately 273 nA at the isocentre (1.7 × 1012 protons s-1) and an estimated peak on-axis average dose rate of 2.0 × 103 Gy s-1. An 11 × 11 spot field delivered 25 Gy in approximately 0.6 s, corresponding to an average field dose rate of 42 Gy s-1, with up to 90 Gy s-1 measured at higher doses. Proton range and Bragg peak width remained unchanged. Spot position deviations were below 0.5 mm, and uniform 4 × 4 cm2 fields with <3% flatness were achieved. CONCLUSIONS: Time-driven delivery enables experimental proton PBS operation at UHDR while maintaining beam characteristics suitable for reproducible dosimetry and radiobiological studies, providing a practical and experimentally characterised platform for preclinical proton PBS FLASH research.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In silico feasibility and robustness evaluation of a single-spot proton Bragg peak FLASH platform for small-animal irradiation using a generalized 3D-printed bolus.Physics in medicine and biology · 2026
- Expanding dose and dose-rate capabilities of a synchrotron-based pre-clinical proton FLASH irradiation platform.Medical physics · 2026
- Single‐field‐uniform‐dose‐per‐fraction simultaneous dose and dose rate optimization (SFUDPF–SDDRO) method for proton FLASH therapyMedical Physics · 2026
- Pulse-resolved spatial dosimetry of ultra-high dose rate electron FLASH beams using optical imaging.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Dynamic Range Modulation via Integrated Universal Range Shifter and Ridge Filter: Enabling Simplified Delivery for FLASH-Enabled IMPT.Physics in medicine and biology · 2026
- Machine Model-Specific Delivery Sequence Optimization for Spot-scanning Proton Arc Therapy Using a Compact Superconducting Synchrocyclotron.Medical physics · 2026