Effects of sparse array geometry and element arrangement on low-intensity transcranial focused ultrasound.
Geunho Shim, Juyoung Park, Kisoo Kim
Department of Biomedical Engineering, College of Electronics and Information, Kyung Hee University, Yongin-si, Gyeonggi-do, Republic of Korea. · Clinical Technology Division, Neumous Inc., Seoul, Republic of Korea.
Key points
From the paper's conclusion- Sparse arrays are used in low-intensity transcranial focused ultrasound, but the effects of element arrangement, aperture and coverage are not well defined.
Abstract
Sparse arrays are used in low-intensity transcranial focused ultrasound, but the effects of element arrangement, aperture and coverage are not well defined. We simulated nine spherical-cap arrays at 250 kHz, combining 50, 60 and 70 mm aperture radii with 40, 50 and 60% coverage (active-element fraction of the cap area), using 8 mm elements, a 100 mm radius of curvature and a fixed nominal drive of 15 W. A scalar hybrid angular spectrum model with straight-ray phase correction was applied to three skull models and three target placements, with lateral steering up to 30 mm. Matched regular arrays, maximum-density regular arrays and added phase-error conditions were included as controls. Increasing coverage produced target-pressure gains close to the square-root scaling expected from element count, with all nine skull-target pairs remaining within 5% of this expectation. Increasing aperture produced smaller foci but less consistent pressure gains. The effect of quasi-random placement, implemented by thinning a regular candidate lattice, was conditional. At 40% coverage and 30 mm steering, quasi-random arrays reduced peak sidelobe ratios by 1.24-5.34 dB relative to matched regular arrays, whereas regular arrays mostly had similar or lower ratios at offsets up to 20 mm. The sidelobe advantage diminished as coverage increased and the regular pitch approached the candidate-lattice pitch. It also depended on the main-lobe exclusion threshold and, at the midbrain target, did not persist under the larger added phase error. Without steering, quasi-random placement increased integrated skull absorption by 10.0-32.9% at 40% coverage. Steering retention varied more strongly across skull-target pairs than across aperture or coverage, whereas focal fragmentation and off-target maxima occurred mostly at the 50 mm aperture. Within the tested design family, arrangement, aperture and coverage should be considered together when selecting arrays for specific steering and focusing requirements. Biological efficacy and thermal safety were not evaluated.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental Validation of Skull Acoustic Modelling Strategies for Transcranial Focused Ultrasound Simulation: A Cross-Comparison Study.Physics in medicine and biology · 2026
- Calibrated simulations for dynamic focusing of ultrasound through the temporal window.Cell reports methods · 2026
- Adaptive Positioning and Targeting Optimization in Transcranial Focused Ultrasound (tFUS): A Numerical Study.Ultrasound in medicine & biology · 2026
- Subwavelength registration of skull-compensating acoustic holograms via nonlinear parametric array feedbackarXiv · 2026
- Computational Modeling of the Acoustic Radiation Force and Viscoelastic Displacement for Transcranial Focused Ultrasound NeuromodulationmedRxiv · 2026
- A deep‐learning model for one‐shot transcranial ultrasound simulation and phase aberration correctionMedical Physics · 2025