Experimental feasibility study of proton therapy range verification by a digital tracking calorimeter prototype.
Alexander Schilling, Johan Alme, Gergely Gábor Barnaföldi, Gábor Bíró, Tea Bodova, Illia Borshchov, Vyacheslav Borshchov, Anthony van den Brink,
Chair for Scientific Computing, RPTU University Kaiserslautern-Landau, Paul-Ehrlich-Straße 36, Kaiserslautern, 67663, Germany. · Department of Physics and Technology, University of Bergen, Allégaten 55, Bergen, Hordaland, 5007, Norway.
Key points
From the paper's conclusion- We find that the aluminum used as absorbers between sensitive layers in the calorimeter serves as a converter for prompt gammas and neutrons.
Abstract
Objective.Uncertainties in the proton range remain an important challenge in proton therapy. In-situ range verification techniques can be used to ensure the Bragg peak depth in the patient matches the treatment plan. Various such methods are under investigation, utilizing the signal from secondary radiation caused by proton interactions with the patient. This study experimentally investigates the feasibility of using a silicon-pixel-based sampling calorimeter for range verification through the detection of secondary photons and neutrons. Approach.The proton range verification capability of a sampling calorimeter based on monolithic active pixel sensors was tested both theoretically and experimentally. We theoretically investigate the plausibility of the planned sampling calorimeter to pick up a signal from the neutral secondary particles distal to the patient. Then, a smaller prototype pixel telescope was constructed to verify our theoretical findings experimentally by placing it behind a water tank irradiated by a medical proton beam with various energies in the range of 100-200 MeV at the Danish Centre for Particle Therapy, Aarhus, Denmark. Main results.We find that the aluminum used as absorbers between sensitive layers in the calorimeter serves as a converter for prompt gammas and neutrons. The combined detection efficiency ranges between 5.8·10-5and 2.7·10-3detector hits per primary proton for beam energies between 50 and 200 MeV. Irradiation of the water phantom leads to a signal in our detector, which is well-correlated with the nominal beam energy. A simulation-based model using only the number of hits per proton can predict the beam energy, acting as a proxy for proton range in a homogeneous target, down to a root mean squared error of 1.34 MeV. Significance.This study marks a promising step towards utilizing a tracking calorimeter designed for proton computed tomography for the secondary application of proton therapy range verification without additional hardware costs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Estimation of dose and linear energy transfer for multi-spot single-energy proton probing beams using in-beam PET and neural network.Physics in medicine and biology · 2026
- Patient-Specific Stopping Power Calibration of Treatment Planning X-Ray CT Using Integrated Depth-Dose Profiles Acquired by Proton Transmission Imaging - A Feasibility Study.Physics in medicine and biology · 2026
- Optimization of selected SPECT system parameters for proton range verification: a Monte Carlo study.Biomedical physics & engineering express · 2026
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026
- Feasibility study of a new type of particle tracker for beam range determination by observation of charged secondary particles emitted perpendicularly from a carbon hadrontherapy beam at energies from 140 to 430 MeV/u.Zeitschrift fur medizinische Physik · 2026
- Experimental verification of clinical dose and LET maps for LET-informed proton therapy.Physics in medicine and biology · 2026