Optimization of selected SPECT system parameters for proton range verification: a Monte Carlo study.
Sharareh Gharib, Payvand Taherparvar
Department of Physics, University of Guilan, University of Guilan, Rasht, 4193833697, Iran (The Islamic Republic of). · Physics, University of Guilan, Namjoo Avenue, Rasht, 41635-1914, Iran (The Islamic Republic of).
Key points
From the paper's conclusion- Overall, appropriate window and septal thickness provided PG localization accuracy of 1-2 mm.
Abstract
This Monte Carlo study investigated a dual head SPECT based system for prompt gamma (PG) range verification in proton therapy, focusing on energy window and collimator septal thickness effects on PG emission profiles and distal dose fall off. Using GATE, correlations between PG emissions (4.44, 5.24, 6.13 MeV) and proton depth-dose distributions were evaluated for 90-180 MeV beams in a PMMA phantom. Narrow (4.2-4.6, 5.0-5.4, 5.9-6.3 MeV) and broad (4.2-6.3 MeV) windows were studied. A dual head SPECT with parallel hole collimator (1-5 mm septa) was simulated. Image reconstruction used FBP and OSEM (STIR 6.2) with attenuation correction. The 5.9-6.3 MeV window showed the highest spatial correlation with proton range (Δ ≈ 1 mm), but its low yield limited applicability. The 4.2-4.6 and 4.2-6.3 MeV windows balanced detection yield and accuracy (Δ ≈ 2-3 mm). Septal thicknesses ≥4 mm produced shadowing artifacts, whereas 3 mm gave the most accurate PG localization for the 90 MeV proton beam studied. For the 4.2-4.6 and 4.2-6.3 MeV windows, the PG localization errors (relative to the true PG emission profile) were 1.5 ± 0.5 and 1.4 ± 0.4 mm with FBP, and 1.9 ± 0.3 and 1.7 ± 0.4 mm with OSEM, despite ~20% lower PG yield than the 1 mm septum. Overall, appropriate window and septal thickness provided PG localization accuracy of 1-2 mm. A 3 mm septum with either the 4.2-4.6 or 4.2-6.3 MeV window offers a favorable trade off between spatial accuracy and detection efficiency under the 90 MeV condition investigated here, highlighting the potential of optimized SPECT systems for proton range monitoring at this energy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026
- Multi-criteria proton treatment planning incorporating spot boosting for prompt gamma in-vivo range verification.Physics in medicine and biology · 2026
- Experimental Assessment of Prompt Gamma Time Imaging in Complex AnatomyarXiv · 2026
- Angular Distributions of Secondary Neutron and Photon Ambient Dose Equivalent in Proton Therapy: A Systematic PHITS 3.35 Study at 70-250 MeV.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Experimental feasibility study of proton therapy range verification by a digital tracking calorimeter prototype.Physics in medicine and biology · 2026
- easyPET.3D Measurement of Positron Emitter Maps in Proton-Irradiated PMMA Phantoms.Biomedical physics & engineering express · 2026