A convergent electron beam approach for small-field irradiation with reduced bremsstrahlung.
Yuma Hayashi, Atsushi Myojoyama
Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University - Arakawa Campus, 7-2-10, Higashiogu, Arakawa-ku, Tokyo, Arakawa, Tokyo, 116-8551, Japan.
Key points
From the paper's conclusion- Bremsstrahlung contamination increased to 2.56%, 1.70%, 1.72%, and 2.11%, respectively, but remained lower than that of the 2 cm tubular applicator at 12 MeV and above.

Significance: Converging electron beams can mitigate fluence loss while reducing bremsstrahlung contamination in small fields.
Abstract
To investigate whether converging electron beams using quadrupole electromagnets can mitigate fluence loss and bremsstrahlung contamination in small irradiation fields. Approach: A three-stage quadrupole electromagnet system was installed beneath a linear accelerator head to converge 6, 9, 12, and 15 MeV electron beams. Dose distributions, dose rate, and bremsstrahlung contamination were evaluated for 2 cm and 1 cm fields. The 2 cm convergent field was compared with a 2 cm tubular applicator, whereas the 1 cm convergent field was evaluated at 5.1 and 10 cm air gaps. Monte Carlo simulations at 6 and 9 MeV assessed detector volume-averaging effects and the origins of bremsstrahlung photons. Main results: The dose rate of the 2 cm convergent field was 1.05-2.53 times higher than that of the 2 cm tubular applicator, while bremsstrahlung contamination was reduced by 0.40-1.79 percentage points. The 1 cm convergent field achieved a therapeutic field width ≤1 cm and penumbra ≤0.85 cm at all energies. At a 5.1 cm air gap, the dose rate was comparable to that of a 3 cm tubular applicator at 9 MeV and exceeded that of a 10 × 10 cm² applicator at 12 and 15 MeV. Bremsstrahlung contamination was 1.08%, 0.87%, 1.04%, and 1.32% at 6, 9, 12, and 15 MeV, respectively, and remained lower than that of the 2 cm tubular applicator at all energies. At a 10 cm air gap, output decreased and dose profiles broadened. Bremsstrahlung contamination increased to 2.56%, 1.70%, 1.72%, and 2.11%, respectively, but remained lower than that of the 2 cm tubular applicator at 12 MeV and above. Significance: Converging electron beams can mitigate fluence loss while reducing bremsstrahlung contamination in small fields. This externally mounted system provides proof of concept for high-spatial-resolution electron beam delivery and future electromagnetic spot-scanning applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Estimation of Very High Energy Electron (VHEE) beam parameters using an in-air screenarXiv · 2026
- Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.Frontiers in oncology · 2026
- Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation.Biomedical physics & engineering express · 2026
- VHEE beam parameter extraction from depth-dose and lateral profile measurements: a Monte Carlo parametric study.Radiological physics and technology · 2026
- A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy.Diagnostics (Basel, Switzerland) · 2026
- Magnetically enhanced peak‐to‐valley dose ratio for megavoltage photon minibeam radiotherapyMedical Physics · 2026