Study of a novel inverted collimator concept for prompt-gamma-based range verification in proton therapy using a hybrid simulation approach.
Korbinian Urban, Giacomo Borghi, Marco Carminati, Andrea Forgione, Martina Piroddi, Christian Riboldi, Letizia Santini, Emma Sofia Bellotti,
Department of Electronics, Information and Bioengineering (DEIB), Politecnico di Milano, Via Giuseppe Ponzio 34, Milan, 20133, Italy. · Department of Energy (DENG), Politecnico di Milano, Via La Masa 34, Milan, 20156, Italy.
Key points
From the paper's conclusion- In addition, we present a first experimental validation of the INV collimator concept at the CNAO hadron therapy facility in Pavia, Italy, demonstrating agreement between the hybrid simulation and measurements.

Significance
These results suggest that the INV collimator is a promising concept for future prompt-gamma cameras, offering competitive performance and facilitating clinical im
Abstract
Objective Prompt-gamma imaging is a promising technique for in-situ range verification during proton therapy, with the potential to reduce range uncertainties. Current prompt-gamma cameras typically rely on mechanical collimators, among which the knife-edge slit (KES) and multi-parallel slit (MPS) collimators are the two most widely studied concepts. In this work, we propose a novel inverted (INV) collimator concept that uses a single bar of high-density material to cast a gamma-ray shadow onto the central region of a pixelated scintillation detector. In addition to advantageous properties such as a high efficiency and a large field of view, the INV collimator also features a significantly reduced weight compared to the KES and MPS collimators. Approach We optimized and compared the different collimator designs with respect to their simulated statistical range retrieval precision. To overcome the computational time limitations associated with full Monte Carlo simulations, we developed a new hybrid simulation tool based on a dedicated fixed-ray calculation of the collimator combined with precalculated Monte Carlo simulations of prompt-gamma emission and detection. This approach enables rapid parameter optimization while incorporating all experimentally relevant effects, such as the detector spectral response and an empirical neutron background. Main results We present optimized collimator geometries for the MPS, KES, and the novel INV collimator. The INV collimator achieves a statistical range retrieval precision (2σ) of 2.2 mm at a proton beam energy of 100 MeV and 2.4 mm at 150 MeV for 108protons irradiating a homogeneous tissue-equivalent phantom, demonstrating performance comparable to the optimized KES design. In addition, we present a first experimental validation of the INV collimator concept at the CNAO hadron therapy facility in Pavia, Italy, demonstrating agreement between the hybrid simulation and measurements. Significance These results suggest that the INV collimator is a promising concept for future prompt-gamma cameras, offering competitive performance and facilitating clinical implementation due to its reduced device weight.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026
- Hierarchical 4π Monte Carlo-based proton beam angle pre-screening framework: a proof-of-concept.Physics in medicine and biology · 2026
- Experimental Assessment of Prompt Gamma Time Imaging in Complex AnatomyarXiv · 2026
- Multi-criteria proton treatment planning incorporating spot boosting for prompt gamma in-vivo range verification.Physics in medicine and biology · 2026
- Optimization of selected SPECT system parameters for proton range verification: a Monte Carlo study.Biomedical physics & engineering express · 2026
- Experimental feasibility study of proton therapy range verification by a digital tracking calorimeter prototype.Physics in medicine and biology · 2026