Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regression
Khaled Ben Abdessalem
Key points
From the paper's conclusion- We propose a robust interval-only synchronization pipeline comprising: (i) Dynamic Repair of abnormally short split intervals caused by extra (false) peaks; (ii) greedy peak-failure-aware mapping with q ∈ { 1 , 2 } to represent isolated missed beats as merged intervals; and (iii) Huber-iteratively reweighted least squares regression with median absolute deviation-based residual gating.
Abstract
Cooperative sensing in clinical instruments, distributed medical devices, and body-area networks requires accurate relative timing between independently clocked sensors, such as electrocardiography (ECG) and photoplethysmography (PPG). Interval-only synchronization operates directly on inter-beat intervals and avoids waveform alignment, but conventional estimators generally assume one-to-one reference–device beat correspondence. Missed beats, motion-induced false peaks, and gross interval errors violate this assumption by creating merged or split intervals and index shifts. We propose a robust interval-only synchronization pipeline comprising: (i) Dynamic Repair of abnormally short split intervals caused by extra (false) peaks; (ii) greedy peak-failure-aware mapping with q ∈ { 1 , 2 } to represent isolated missed beats as merged intervals; and (iii) Huber-iteratively reweighted least squares regression with median absolute deviation-based residual gating. Approximate uncertainty is quantified using an OLS-style Newey–West HAC covariance estimate (approximate and conditional when gating) and Fieller confidence sets for the ratio-defined delay-dynamics parameter. Monte Carlo ablation on synthetic ECG/PPG data (ORACLE and SIGNAL pipelines) shows that structural peak-detection failures can be catastrophic for OLS (e.g. RMS E after , all ≈ 227 ms under Extra 10% and ≈ 232 ms under a mixed extreme regime). In contrast, the proposed pipeline maintains robustness; in the mixed extreme regime it achieves RMS E after , in ≈ 37.4 ms (SIGNAL) and ≈35.9 ms (ORACLE). The real-data experiment was performed as a semi-synthetic feasibility validation using matched ECG/PPG beats from a real recording, with known clock skew imposed on the PPG timestamps and controlled peak failures injected into the resulting device interval sequence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring SystemsCardiovascular Engineering and Technology · 2026
- Recording sampling rate as a negative-control exposure for detecting measurement-driven bias in heart-rate-variability-outcome associations: a quantitative bias analysis across ICU waveform databasesmedRxiv · 2026
- MATH-SYNC: Robust and Identifiable Affine-Clock Alignment of Heterogeneous Multirate Sensor StreamsMathematics · 2026
- Robust PPG-Exclusive Heart Rate Estimation During Maximal Exercise Using Population-Based PriorsarXiv · 2026
- Multi-Sensor Fusion for False-Alarm Reduction by Timed Electromechanical CorroborationarXiv · 2026
- ARTA: Graph Combinatorial Optimization for Robust Interbeat Interval Estimation from Wearable Electrocardiogram SignalsarXiv · 2026