Count-nonlinearity correction of single-photon avalanche diodes using laser superposition and reference-photon-rate validation
wentao Wang, Qing Kang, Jianjun Li, Xiaobing Zheng
Key points
From the paper's conclusion- We present a measurement-based correction framework combining continuous-wave laser-superposition linearity evaluation, a wavelength-shared rate-dependent effective-dead-time model, and independent reference-photon-rate validation using a silicon trap detector.
Abstract
Single-photon avalanche diode (SPAD) count outputs exhibit rate-dependent compression at high count rates, limiting their linear operating range in weak-light radiometry. We present a measurement-based correction framework combining continuous-wave laser-superposition linearity evaluation, a wavelength-shared rate-dependent effective-dead-time model, and independent reference-photon-rate validation using a silicon trap detector. The model was identified from 33 superposition groups at 460 nm, 532 nm, and 633 nm, covering approximately 10^3–2 × 10^6 s−1, and was frozen before independent validation. After correction, the RMS superposition residuals were 0.114%, 0.280%, and 0.293%, respectively, with a pooled RMS of 0.243%; all 33 groups satisfied the ±1% linearity criterion. Odd–even and leave-one-wavelength-out tests yielded test RMS values below 0.295%. Independent validation using 78 separately acquired points extending to 2.5 × 10^6 s−1 reduced zero-intercept fitting RMS residuals from 5.811%–6.157% to 0.232%–0.336%. Overall, 77 of 78 points satisfied the ±1% criterion, while all 30 points at or above 1.0 × 10^6 s−1 satisfied the ±0.5% criterion. These results demonstrate substantially improved SPAD count linearity within the validated continuous-wave operating range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Practical Calibration Method for Correcting Count-Rate Nonlinearity in Free-Running Silicon SPADsPhotonics · 2026
- Self-Referenced Diagnosis of SPAD Dark-Count-Rate Switching for Quantum Communication ReceiversarXiv · 2026
- Fast SPAD-array timing-error correction with time-referencing for non-line-of-sight imaging.Optics express · 2026
- Optical ISAC with Single-Photon Avalanche Diode (SPAD) ArraysarXiv · 2026
- A Triple Compensation Method for Correcting Temperature-Induced Distance Errors in SPAD-Based d-ToF LiDARElectronics · 2026
- A Fully Circular P+/N-Well CMOS Diode with Low Dark Count Rate for Single-Photon Imaging.Sensors (Basel, Switzerland) · 2026