From the paper's conclusion
  • This work showed that the biomechanical performance of implant thread geometry is strongly dependent on the host bone quality.

Abstract

This study investigated the influence of dental implant thread geometry on bone remodeling using finite element analysis coupled with a remodeling algorithm. Four thread designs (square, buttress, reverse-buttress, and triangular) were evaluated in D2 and D4 bone, using 2D axisymmetric models. It was shown that in D2 there is almost the same extent of bone density reduction in all four thread shapes. But, in D4, first, an increase in bone density was predicted due to remodeling, second significant differences were observed for different thread shapes. In D4 bone type, the square- and buttress threads showed more favorable long-term remodeling responses, i.e.. greater increase in apparent density and Young's modulus of elasticity. This work showed that the biomechanical performance of implant thread geometry is strongly dependent on the host bone quality. In vivo studies, and long-term clinical follow up are needed to validate the theoretical predictions made here.