Validity and reliability of the K5 wearable metabolic system during high-intensity treadmill testing in healthy, young adults
Katelyn D. Brown, Belinda Sanchez, Monica Bennett, Jakob L. Vingren, Taylor Gilliland, Megan Reynolds, Alan Jones, Simon Driver
Baylor Scott & White Health · University of North Texas
Key points
From the paper's conclusion- This study evaluated the validity and test-retest reliability of K5 relative to the laboratory-based ParvoMedics TrueOne2400® system (PM) during treadmill cardiopulmonary exercise testing (CPET) in healthy young adults.Methods Thirty-one adults (mean age 25.8 ± 2.8 years, 61.3% male) completed three Bruce protocol CPET trials: one using PM and two using K5.
Abstract
Background The COSMED K5 portable metabolic system (K5) enables field-based assessments of gas exchange variables. This study evaluated the validity and test-retest reliability of K5 relative to the laboratory-based ParvoMedics TrueOne2400® system (PM) during treadmill cardiopulmonary exercise testing (CPET) in healthy young adults.Methods Thirty-one adults (mean age 25.8 ± 2.8 years, 61.3% male) completed three Bruce protocol CPET trials: one using PM and two using K5. The order of the first two trials was randomized and counter balanced; the third used K5. Breath-by-breath data were averaged over the final 90 seconds of each CPET stage. Validity was assessed using paired comparisons and Bland-Altman analyses. Reliability was evaluated using intraclass correlation coefficients (ICC).Results The K5 demonstrated excellent test-retest reliability for oxygen consumption (VO2) and carbon dioxide production (VCO2) (ICC = 0.90–0.98). Relative to PM, K5 values for VO2 and VCO2 were slightly higher at lower intensities and lower near peak intensities (P < 0.01). Bland-Altman analyses revealed proportional bias for VO2 and VCO2 across workloads.Conclusions The K5 is a valid and reliable device for treadmill CPET in healthy young adults. Small systematic differences exist relative to PM at extreme workloads yet remain within acceptable clinical limits, supporting its utility for longitudinal and applied assessments requiring portability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validity and reliability of the Vyntus CPX metabolic gas analyzer compared with the Douglas Bag method.European journal of applied physiology · 2026
- The accuracy of VO2 max, heart rate and energy expenditure measurements from Google Pixel Watch 3.PloS one · 2026
- Physiological Associations, Within-Session Repeatability, and Exploratory Prediction of Peak Oxygen Uptake Using the Revised 3-Minute Seated Stepping Test in Healthy Young Japanese Adults.Journal of functional morphology and kinesiology · 2026
- Agreement of four metabolic equations for estimating submaximal cycling oxygen uptake with two portable indirect calorimetry systems in untrained young women.BMC sports science, medicine & rehabilitation · 2026
- Cardiopulmonary exercise testing (CPET) in an elite, peri-menopausal female runner using Bruce protocol: A case report.Medical journal, Armed Forces India · 2026
- ERS technical standard on reference values for cardiopulmonary exercise testing: summary report and a call for actionEuropean Respiratory Journal · 2026