Validity and reliability of the Vyntus CPX metabolic gas analyzer compared with the Douglas Bag method.
Kim Arne Heitmann, Karianne Hagerupsen, Svein Arne Pettersen, Boye Welde, Jon Ingulf Medbø, Kristoffer Robin Johansen
School of Sport Sciences, UiT The Arctic University of Norway, Post Box 6050 Langnes, N-9037, Tromsø, Norway. kim.a.heitmann@uit.no. · School of Sport Sciences, UiT The Arctic University of Norway, Post Box 6050 Langnes, N-9037, Tromsø, Norway.
Key points
From the paper's conclusion- The Vyntus CPX demonstrated agreement comparable with other studies on similar commercial instrument but showed reduced accuracy at higher levels of V̇O2 and V̇CO2. The variability in technological error was somewhat larger than what is generally considered acceptable for gas exchange measurements. The Vyntus CPX showed systematic test-retest differences for V̇CO2 and RER, with proportional bias evident for all measured gas exchange variables.
Abstract
AIMS: To validate the Vyntus CPX system against Douglas Bag method. METHODS: Twelve cycle-trained men and women completed an incremental cycling protocol starting at 50 W, increasing by 50 W every fifth minute up to 350 W or to near exhaustion. After a 5 min break a test of peak oxygen uptake (V̇O2peak) was carried out. Gas exchange parameters were continuously measured using the Vyntus CPX system in mixing chamber mode, while expired air was simultaneously collected in Douglas Bags during the final 60 s of each stage. RESULTS: Mean systematic differences were observed for V̇O2 (+ 0.108 L/min, 4.6%) and respiratory exchange ratio (RER, -0.032, -3.2%). The Vyntus CPX measured proportionally higher V̇O2 (+ 0.032 L/min) and lower V̇CO2 (-0.032 L/min) per L/min increase in V̇O2 and V̇CO2, respectively, measured by the Douglas Bag. Consequently, the Vyntus CPX proportionally underestimated RER by 0.017 per 0.1-unit increase. From paired measurements on two days, coefficient of variation (CV) for the gas exchange variables were 3.4%-5.7%. After adjusting for biological variability, systematic test-retest differences were observed for V̇CO2 (+ 0.066 L/min, 2.0%) and RER (+ 0.021, 2.3%). Proportional bias was evident across all gas exchange variables. CONCLUSION: The Vyntus CPX demonstrated agreement comparable with other studies on similar commercial instrument but showed reduced accuracy at higher levels of V̇O2 and V̇CO2. The variability in technological error was somewhat larger than what is generally considered acceptable for gas exchange measurements. The Vyntus CPX showed systematic test-retest differences for V̇CO2 and RER, with proportional bias evident for all measured gas exchange variables.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement of four metabolic equations for estimating submaximal cycling oxygen uptake with two portable indirect calorimetry systems in untrained young women.BMC sports science, medicine & rehabilitation · 2026
- Validity and reliability of the K5 wearable metabolic system during high-intensity treadmill testing in healthy, young adultsBaylor University Medical Center Proceedings · 2026
- A Comparison of Different Models to Compute Cardiac Output by Open-Circuit Inert Gas Wash-In.Medicine and science in sports and exercise · 2026
- Modified automated detection of the first ventilatory threshold using V-slope regression: validation against manual assessment in healthy adultsEuropean Heart Journal · 2026
- ERS technical standard on reference values for cardiopulmonary exercise testing: summary report and a call for actionEuropean Respiratory Journal · 2026
- Exercise intensity alters the relationship between breath acetone output and oxygen uptake during cycling exercise: a comparison of submaximal and incremental maximal protocols.Journal of breath research · 2026