A Laboratory Investigation of Medication Delivery by Pressurized Metered Dose Inhaler with Valved holding chamber Using a Model Adult Oropharyngeal Airway and Inhalation Flow Profiles from Trained Healthy Adults.
Mark W Nagel, Jason A Suggett, Jolyon P Mitchell
Trudell Medical International, 725 Baransway, London, Ontario, Canada N5V 5G4. · Jolyon Mitchell Inhaler Consulting Services (JMICS) Inc., 1154 St. Anthony Road, London, Ontario, Canada, N6H 2R1.
Key points
From the paper's conclusion- We report a laboratory-based study measuring how inhalation profiles from trained adults may affect inhaled potential lung-delivered mass of a widely prescribed short-acting bronchodilator via a pMDI-valved holding chamber (spacer) combination.
Abstract
Patient inhalation flow-elapsed time profiles using pressurized metered dose inhalers (pMDIs) can be highly variable for several reasons, including inability to maintain a consistent flow rate-time profile. We report a laboratory-based study measuring how inhalation profiles from trained adults may affect inhaled potential lung-delivered mass of a widely prescribed short-acting bronchodilator via a pMDI-valved holding chamber (spacer) combination. Adult inhalation profiles from a previous study inhaling room air without medication were obtained after (a) taking a long slow inhalation followed by a 5-s breath hold (method A); (b) inhaling tidally for 2-3 respiration cycles (method B). Each profile was replayed in a simulation where albuterol (100 μg/actuation) was 'inhaled' in vitro via an adult oropharyngeal model connected to a filter distal to the model, with flow controlled via breathing simulator. Filter-collected albuterol provided total mass/actuation available for lung delivery (3-replicates/individual/inhalation profile). Total inhaled volume varied widely and ranges between sexes differed (M: 155-5002 ml, F:214-2711 ml - deep inhalation; M: 617-7722 ml, F:450-3789 ml - tidal breathing). Large variability was observed in potential lung-deliverable drug mass (RSD = 30.1% method A; 30.3% method B) across all inhalation profiles for either method. Intra-profile variability was no worse with either breathing maneuver. Large inter-subject variability in inhalation profiles also existed with either method A or method B. The primary function of the valved holding chamber to retain medication until inhaled was met, since some albuterol was delivered to the filter distal to the oropharyngeal model even in the worst cases.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Changes in device-matched peak inspiratory flow after bedside inhaler coaching and factors associated with residual suboptimal flow: a retrospective observational study.Frontiers in pharmacology · 2026
- Enhancing aerosol delivery in asthma and COPD: a comparison of MDI, valved holding chamber, and DPI systems using functional respiratory imaging (FRI)Scientific Reports · 2026
- Experimental and computational study of pMDI aerosol delivery in mucus-coated pediatric airways: Effects of inhalation flow and actuation timing.Journal of pharmaceutical sciences · 2026
- Translational feasibility of acoustic monitoring for cumulative lung-delivered dose during routine dry powder inhaler use.International journal of pharmaceutics · 2026
- Peak Inspiratory Flow in COPD Patients Treated with Pressurized Metered-Dose Inhalers and Spacers: Implications for Inhaler Selection. A Multicenter Cross-Sectional Study.International journal of chronic obstructive pulmonary disease · 2026
- Modeling exhaled aerosol deposition in an impaction filter across clinically measured breath flowrates.Journal of breath research · 2026