Peak Inspiratory Flow in COPD Patients Treated with Pressurized Metered-Dose Inhalers and Spacers: Implications for Inhaler Selection. A Multicenter Cross-Sectional Study.
Lara Bravo Quiroga, Jose Miguel González-Moro Rodríguez, Francisco Javier Álvarez-Gutiérrez, Krasimira Baynova, Mariam de la Poza Abad, Jose Luis Izquierdo-Alonso
Department of Pneumology, Hospital Universitario Príncipe de Asturias, Alcalá de Henares, Madrid, Spain. · Department of Pneumology, Hospital Universitario Virgen del Rocío, Sevilla, Spain.
Key points
From the paper's conclusion- Most COPD patients treated with pMDIs and spacers appear to have sufficient inspiratory capacity to use medium- to high-resistance DPIs. Routine objective assessment of PIF may help optimize individualized inhaler selection and identify patients who could benefit from alternative inhaler devices.
Abstract
BACKGROUND: Spacer devices are frequently prescribed with pressurized metered-dose inhalers (pMDIs) in patients with COPD for a variety of clinical and practical reasons, including concerns about inspiratory capacity, inhaler technique, and hand-breath coordination. However, inspiratory flow is rarely measured objectively. PURPOSE: To determine whether COPD patients treated with pMDIs and spacers can generate adequate peak inspiratory flow (PIF) for dry powder inhaler (DPI) use, to evaluate the impact of incomplete exhalation, and to examine patient perceptions of spacer use. PATIENTS AND METHODS: This multicenter, cross-sectional study included 273 clinically stable COPD patients treated with pMDIs and spacers across 70 Spanish centers. PIF was measured with the In-Check Dial at medium-low (R2) and medium-high (R4) resistance under both standard and suboptimal conditions. Patient knowledge, satisfaction, and perceived barriers were evaluated using structured study questionnaires. RESULTS: Overall, 96.0% of patients achieved a PIF ≥30 L/min at R4, whereas 61.9% reached ≥60 L/min. Incomplete exhalation had only a limited effect on inspiratory performance, with 94.1% of patients maintaining a PIF ≥30 L/min at R4. Lower PIF was associated with greater dyspnea and more severe airflow limitation in bivariate analyses. Spacer prescription was seldom based on objective PIF assessment. Most patients reported practical limitations associated with spacer use and expressed a preference for simpler inhaler devices. CONCLUSION: Most COPD patients treated with pMDIs and spacers appear to have sufficient inspiratory capacity to use medium- to high-resistance DPIs. Routine objective assessment of PIF may help optimize individualized inhaler selection and identify patients who could benefit from alternative inhaler devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Changes in device-matched peak inspiratory flow after bedside inhaler coaching and factors associated with residual suboptimal flow: a retrospective observational study.Frontiers in pharmacology · 2026
- Suboptimal peak inspiratory flow, COPD exacerbations, hospitalisations and disease burden: a longitudinal follow-up in a multicentre real-world study.ERJ open research · 2026
- The Impact of Inspiratory Muscle Function on Inhaler Device Adaptability in Stable COPD Patients.Journal of aerosol medicine and pulmonary drug delivery · 2026
- Enhancing aerosol delivery in asthma and COPD: a comparison of MDI, valved holding chamber, and DPI systems using functional respiratory imaging (FRI)Scientific Reports · 2026
- Personalizing Inhaled Therapy in COPD and Asthma: Integrating Drug, Device, and Patient Factors.Biomedicines · 2026
- Bronchodilator Drug Delivery With Pressurized Meter-Dose-Inhaler Versus Jet Nebulizer With Noninvasive-Ventilation in Acute-Exacerbation of COPD: A Randomized Controlled Pharmacokinetic Endpoint Trial.Respirology (Carlton, Vic.) · 2026