The Review of scientific instruments

Experimental setup for testing nanocalorimeter sensors as a plasma diagnostics tool.

Carles Corbella, Feng Yi, Andrei Kolmakov

Materials Measurement Science Division, MML, NIST, Gaithersburg, Maryland 20899, USA. · Nanoscale Device Characterization Division, PML, NIST, Gaithersburg, Maryland 20899, USA.

Review

Key points

From the paper's conclusion
  • Here, we describe the experimental setup for complementary characterization of nanocalorimeters’ performance using plasma diagnostics techniques.

Abstract

The application of microfabricated nanocalorimeters for real-time, highly sensitive, and selective monitoring of the energy fluxes from low-pressure plasma discharges has been reported recently. Here, we describe the experimental setup for complementary characterization of nanocalorimeters' performance using plasma diagnostics techniques. The Langmuir probe, retarding field energy analyzer, optical emission spectroscopy, quadrupole mass spectrometry, and thermocouple-based thermometry constitute the plasma monitor suite. We discuss two application examples of the plasma monitor suite: (1) discrimination between electron- and ion-induced heat fluxes via sensor direct current biasing, and (2) detection of fast temperature transients in pulsed discharges. The main goal of this work is to demonstrate a nanocalorimetry platform as a real-time metrology for plasma processing and process control.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Experimental setup for testing nanocalorimeter sensors as a plasma diagnostics tool — Science Skim