The Review of scientific instruments

Collection, characterization, and precision measurement of levitated charged nanoparticles.

B E Kane, Joyce Coppock, Sunghyun Kim, Sarah Westgate

Laboratory for Physical Sciences, 8050 Greenmead Dr., College Park, Maryland 20740, USA. · University of Maryland, College Park, Maryland 20742, USA.

Key points

From the paper's conclusion
  • We discuss methods for particle generation and collection using electrospray emission, rapid characterization by direct imaging of thermal motion, and transfer of the particle from the trap where it is collected to a separate analysis trap in order to achieve better vacuum and lower noise.

Abstract

We describe apparatus and experimental procedures for high stability precision measurements of levitated nanoscale particles confined in an ion trap in high vacuum. We discuss methods for particle generation and collection using electrospray emission, rapid characterization by direct imaging of thermal motion, and transfer of the particle from the trap where it is collected to a separate analysis trap in order to achieve better vacuum and lower noise. In the analysis trap at high vacuum (pressure p ≃ 10-8 Torr), we employ thermostatic control of the trapped particle oscillation amplitudes, allowing long-term, precision measurements of oscillation frequencies, from which the charge to mass ratio (Q/M) can be deduced. Under these conditions, we achieve Q/M measurement precision approaching 10-5. This sensitivity will enable, for example, investigations of the surface chemistry of μm-scale levitated materials in ultra-high vacuum environments.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Collection, characterization, and precision measurement of levitated charged nanoparticles — Science Skim