CFD-based dynamic performance investigation of a cylindrical gear flowmeter under step and pulsating flow
Xinyue Zhao, Hongyang Ren, Yongsheng Zhang, Fan Wang, Suna Guo
Key points
From the paper's conclusion- Under pulsating flow, the measured amplitude decays significantly with rising frequency; the maximum attenuation exceeds 70% within 10–50 Hz, and attenuation is more severe at low flow rates.
Abstract
Cylindrical gear flowmeters are high-precision positive-displacement instruments widely used in aerospace, chemical engineering, and other fields requiring strict dynamic metering. However, research on their dynamic characteristics under fluctuating flow conditions remains insufficient, and conventional static calibration cannot reflect real operating performance. To address these gaps, this paper takes a DN16 cylindrical gear flowmeter as the research object and proposes a computational fluid dynamics simulation method integrated with a rotation (one degree of freedom) model. The time-varying inlet velocity is controlled by function expressions defined in the built-in C-Expression. Simulations are conducted under step flow and pulsating flow conditions, and the method is validated via real-flow experiments. Results indicate that the flowmeter behaves as a first-order inertial system under step flow. The time constant decreases with increasing step amplitude, and the response lag is more pronounced under negative step excitation. Under pulsating flow, the measured amplitude decays significantly with rising frequency; the maximum attenuation exceeds 70% within 10–50 Hz, and attenuation is more severe at low flow rates. The phase difference increases, up to 140°, with increasing frequency. This study provides a technical foundation for dynamic calibration and error compensation of cylindrical gear flowmeters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Characterization of Dynamic Characteristics of Pneumatic Systems Based on Aerodynamic Frequency RatioSAE technical papers on CD-ROM/SAE technical paper series · 2026
- Mass conductance of a low pressure LPG injector as a function of valve lift and nozzle diameterJournal of Physics: Conference Series · 2026
- Geometry-dependent gas–liquid transport, pressure pulsation, and energy dissipation in a π-shaped ultrasonic-flowmeter conduitPhysics of Fluids · 2026
- Quantitative modeling of on-chip peristaltic pump with air-damped microfluidic deviceJournal of Physics: Conference Series · 2026
- Simulation Research on the Effects of Gas Static Mixers Based on Different StructuresApplied Sciences · 2026
- Numerical simulation and experimental investigation of flow resistance performance for downstream duct simulator of civil aircraft mixing chamberJournal of Physics: Conference Series · 2026