Journal of Physics: Conference Series

Quantitative modeling of on-chip peristaltic pump with air-damped microfluidic device

Lei Chen, Xinjie Zhang, Yafei Guo, Zhinan Yang

Key points

From the paper's conclusion
  • To address flow pulsations of on-chip peristaltic pumps that may damage shear-sensitive biological samples, this paper presents a coupled model of an on-chip peristaltic pump integrated with a multi-stage air-damped microfluidic device.

Abstract

To address flow pulsations of on-chip peristaltic pumps that may damage shear-sensitive biological samples, this paper presents a coupled model of an on-chip peristaltic pump integrated with a multi-stage air-damped microfluidic device. First, a time-varying pump source model based on Gaussian functions is proposed to decouple the process of roller compression and fluidic backflow. Second, considering the nonlinear polytropic thermodynamic process of microscale gas compression, a dynamic fluidic capacitance model is established to construct a high-order fluidic resistance-capacitance ( RC ) ladder network. Quantitative analyses show that the fluidic pulsation rate monotonically attenuates with increased damping stages and rotational speed, where the multi-stage cascaded dampers eliminate pulsations rapidly under high rotational speeds. Moreover, increasing damper volume enhances fluidic capacitance but exhibits a prominent diminishing marginal return. This unified model provides a theoretical basis for the active design and parametric optimization of high-performance, low-pulsation microfluidic peristaltic pumps.