The Review of scientific instruments

Probe with four independent detectors for parallel detection in high-resolution solution-state NMR.

Sebastiaan Van Dyck, Coline Wiame, Hadi Loutfi, Anna Sonnefeld, Philippe Pelupessy, Kirill Sheberstov, Sagar Wadhwa, Ronald Kampmann,

Chimie Physique et Chimie du Vivant (CPCV, UMR 8228), Department of Chemistry, Ecole Normale Supérieure, PSL University, Sorbonne University, CNRS, Paris 75005, France. · Voxalytic GmbH, Rosengarten 3, 76228 Karlsruhe, Germany.

Key points

From the paper's conclusion
  • A quadruple probe was designed for wide-bore spectrometers with four independent detectors, with sample volumes of ∼5 μl each.

Abstract

A quadruple probe was designed for wide-bore spectrometers with four independent detectors, with sample volumes of ∼5 μl each. Four transmission-line detectors are equipped with radio frequency channels for 1H (or 19F) and 13C, which can be tuned and matched independently. Each detector is equipped with 6 on-chip shim-coils for local shimming. Linear pulsed field gradients can be applied independently to each sample along the z-axis (in the B0 direction). Slice selection allows one to reduce the effective volumes to ∼2.5 μl to improve the resolution and minimize crosstalk. Four sets of experiments can be acquired simultaneously on four different samples, for example, four inversion-recovery or four Carr-Purcell-Meiboom-Gill experiments.