◆ The Review of scientific instruments2026-08-01

Probe with four independent detectors for parallel detection in high-resolution solution-state NMR.

Sebastiaan Van Dyck, Coline Wiame, Hadi Loutfi, Anna Sonnefeld, Philippe Pelupessy, Kirill Sheberstov, Sagar Wadhwa, Ronald Kampmann, Jan G Korvink, Geoffrey Bodenhausen

原始摘要(英文原文)· Original abstract
A quadruple probe was designed for wide-bore spectrometers with four independent detectors, with sample volumes of ∼5 μl each. Four transmission-line detectors are equipped with radio frequency channels for 1H (or 19F) and 13C, which can be tuned and matched independently. Each detector is equipped with 6 on-chip shim-coils for local shimming. Linear pulsed field gradients can be applied independently to each sample along the z-axis (in the B0 direction). Slice selection allows one to reduce the effective volumes to ∼2.5 μl to improve the resolution and minimize crosstalk. Four sets of experiments can be acquired simultaneously on four different samples, for example, four inversion-recovery or four Carr-Purcell-Meiboom-Gill experiments.