Magnetic resonance imaging

Combined volume transmit/receive and surface receive-only coils for simultaneous brain and spinal cord imaging of squirrel monkey at 9.4 T.

Ming Lu, Feng Wang, Daniel Colvin, Xuanwei Chang, Li Min Chen, John C Gore, Xinqiang Yan

Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN, 37232, USA. Electronic address: ming.lu@vumc.org. · Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN, 37232, USA; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, 37232, USA.

Key points

From the paper's conclusion
  • The combined two-channel coil enables simultaneous anatomical imaging of the brain and cervical spinal cord at 9.4 T using only two receive channels.

Abstract

PURPOSE: A combined volume transmit/receive (T/R) and surface receive-only coil was developed for simultaneous anatomical brain and cervical spinal cord imaging of squirrel monkeys at 9.4 T, for which dedicated hardware is lacking in small-bore high-field scanners. METHODS: An Alderman-Grant resonator (AGR) served as the volume T/R coil, and a co-positioned saddle-shaped surface receive coil increased sensitivity over the cervical spinal cord. The two coils produced orthogonal magnetic fields, giving intrinsic geometric decoupling. Bench characterization included unloaded and loaded quality factors. Signal-to-noise ratio (SNR) was measured by gradient-echo imaging of two phantoms, an ex vivo specimen, and a live squirrel monkey, and compared across four receive configurations: a reference quadrature birdcage coil, the AGR alone, the surface coil alone, and the two combined. RESULTS: Inter-coil isolation was -19.9 dB or better under all loading conditions. In vivo, the combined configuration increased cord SNR 2.35-fold over the AGR alone and 2.02-fold over the birdcage coil, giving a brain-to-cord ratio of 1.02 against 2.76. The ex vivo specimen gave a comparable gain over the birdcage coil, ranging from 1.97-fold to 2.21-fold, whereas the more heavily loading saline phantoms gave up to 3.57-fold. Brain SNR was 0.67 to 0.82 times the birdcage value, a deficit already present with the AGR alone and therefore not caused by the surface channel. CONCLUSION: The combined two-channel coil enables simultaneous anatomical imaging of the brain and cervical spinal cord at 9.4 T using only two receive channels.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers