The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons

Minimally Invasive Distal Metatarsal Osteotomy: Does Screw Design, Number, and Position Matter?

Ali A Khazaal, Nathaniel Helwig, Ashley Thai, Sayud Ali, Ame Anteyi, Noman A Siddiqui

Department of Orthopedic Surgery, Rubin Institute of Advanced Orthopedics, International Center for Limb Lengthening, 2401 West Belvedere Ave. Baltimore, MD 21215. · Department of Orthopedic Surgery, Rubin Institute of Advanced Orthopedics, International Center for Limb Lengthening, 2401 West Belvedere Ave. Baltimore, MD 21215. Electronic address: nsiddiqui@lifebridgehealth.org.

Observational · n=221

Key points

From the paper's conclusion
  • Infection rates did not differ significantly. Compression screws were associated with earlier union, although this association was attenuated after calendar year adjustment. Screw position was correlated with time to union, while additional and compression screws were associated with higher rates of hardware complications.

Abstract

BACKGROUND: Optimal fixation for minimally invasive distal metatarsal osteotomy remains uncertain. STUDY DESIGN: Retrospective cohort study PURPOSE: Examine associations of screw number and design with time to first documented radiographic union and explores screw-position correlations with time to union and postoperative complications. METHODS: Adults undergoing hallux valgus correction from January 2019 to March 2025 were included; prior correction, fixation-free surgery, and inadequate radiographic follow-up were exclusions. RESULTS: Among 221 patients (273 feet), time to union did not differ between screw number (difference, -1.38 weeks; 95% CI, -4.42 to 1.66; p = 0.372). Compression screws were associated with shorter time to union than non-compression screws (difference, -5.01 weeks; 95% CI, -8.48 to -1.54; p = 0.005), including in multivariable analysis (p = 0.012), but not after calendar year adjustment (p = 0.163). Greater osteotomy-to-proximal screw distance correlated with longer time to union (ρ = 0.266; p < 0.001), whereas greater screw head-to-first tarsometatarsal joint distance correlated with shorter union time (ρ = -0.314; p < 0.001). Hardware irritation and removal were more frequent in 2 versus 1-screws (13% vs 1%, p = 0.002; 12% vs 1%, p = 0.004) and compression versus non-compression constructs (16% vs 2%, p < 0.001; 15% vs 2%, p = 0.001). CONCLUSION: Infection rates did not differ significantly. Compression screws were associated with earlier union, although this association was attenuated after calendar year adjustment. Screw position was correlated with time to union, while additional and compression screws were associated with higher rates of hardware complications. LEVEL OF CLINICAL EVIDENCE: Therapeutic Level 3.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Minimally Invasive Distal Metatarsal Osteotomy: Does Screw Design, Number, and Position Matter? — Science Skim