Validation of a dynamic evaporation technique for generating isoflurane/air gas mixtures standards to calibrate anesthetic gas monitors.
Aksam Abdelkhalik, Ghada Makhlouf, Randa N Yamani
Gas Analysis and Fire Safety Laboratory, National Institute of Standards (NIS), El‑Sadat Street, El‑Haram, P.O. Box 136, El‑Giza, 12211, Egypt. aksam.abdelkhalik@nis.sci.eg. · Gas Analysis and Fire Safety Laboratory, National Institute of Standards (NIS), El‑Sadat Street, El‑Haram, P.O. Box 136, El‑Giza, 12211, Egypt.
Key points
From the paper's conclusion- The dynamic evaporation and mixing technique standardized in EN 1839 (method T), originally developed to determine the flammability limits of chemical vapors, was applied to prepare isoflurane-in-air calibration gas mixtures, with air serving as the carrier gas.
Abstract
The dynamic evaporation and mixing technique standardized in EN 1839 (method T), originally developed to determine the flammability limits of chemical vapors, was applied to prepare isoflurane-in-air calibration gas mixtures, with air serving as the carrier gas. Isoflurane concentrations spanning 0 to 8 vol% were generated on demand by independently controlling the liquid isoflurane and air flow rates. Verification using an anesthetic gas monitor demonstrated excellent agreement between predicted and measured concentrations. A comprehensive validation study was conducted in accordance with EURACHEM guidelines, encompassing linearity, limit of detection (LOD), limit of quantification (LOQ), precision, robustness, and measurement uncertainty. The method yielded a correlation coefficient of r² = 0.9998 over the full clinical range, an LOD of 0.08 vol%, and an LOQ of 0.27 vol%. All concentrations between 0.70 and 8 vol% complied with EN 1839 uncertainty requirements, confirming the suitability of this approach for routine calibration of isoflurane gas monitors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of the Content of the Main Components of Breathing Mix. Part IIPolish Hyperbaric Research · 2026
- Washout characteristics of sevoflurane and desflurane in surgical patients: a prospective observational study using gas chromatography - ion mobility spectrometry.BMC anesthesiology · 2026
- End-Tidal Propofol Monitor Based on Deep Ultraviolet Fluorescence Spectroscopy Using Physics-Based Real-Time Correction and Deconvolution ModelsAnalytical Chemistry · 2026
- Combining an anaesthetic conserving device with a circle breathing system to minimise isoflurane waste: an in vitro study.Veterinary anaesthesia and analgesia · 2026
- Assessing inhalation technique: a comparison of three methods in patients with asthma and COPDnpj Primary Care Respiratory Medicine · 2026
- Measurement Validation as Scientific Claim Admissibility: An Evidence-Gated Decision FrameworkMeasurement Science and Technology · 2026