Exploratory proof-of-concept study and failure mode analysis of a dual-lumen stent graft for transapical off-pump LVAD implantation.
Florian Meissner, Manuela Schön, Michelle Costa Galbas, Hendrik Straky, Johannes Dinkelaker, Martin Czerny, Wolfgang Bothe
Department for Cardiac, Thoracic, Vascular Surgery and Aortic Medicine, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Strasse 55, 79106, Freiburg, Germany. · Center for Experimental Models and Transgenic Service, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, Stefan-Meier-Strasse 17, 79104, Freiburg, Germany.
Key points
From the paper's conclusion- This study represents the first exploratory proof-of-concept evaluation of a dual-lumen stent graft designed to redirect blood from the left ventricle across the aortic valve into the ascending aorta while enabling LVAD support.
Abstract
Transapical off-pump implantation of left ventricular assist devices (LVADs) requires a separate inflow cannula and outflow graft. This study represents the first exploratory proof-of-concept evaluation of a dual-lumen stent graft designed to redirect blood from the left ventricle across the aortic valve into the ascending aorta while enabling LVAD support. The stent graft was tested in a mock circulatory loop, during wet-lab experiments on explanted porcine hearts, and in an exploratory acute large-animal study. In vitro testing showed no relevant impairment of pump performance. Wet-lab experiments confirmed that transapical positioning is feasible while revealing incomplete device expansion and myocardial tearing during balloon-assisted apical displacement. In vivo, a reinforced prototype could be positioned across the aortic valve, but myocardial compression impaired inflow expansion and LV unloading, while insufficient apical sealing resulted in bleeding and air embolism. Postmortem analysis confirmed correct device positioning without macroscopic valvular injury but identified myocardial tearing. These failure modes highlight the need for alternative design strategies for transapical off-pump LVAD implantation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Left ventricular assist device implantation in a 'turtle-shell' sclerotic left ventricular apex: a case report of landing-zone selection and tailored fixation strategy.European heart journal. Case reports · 2026
- Axillary microaxial flow pump support for stepwise source control and bridge to durable left ventricular assist device therapy in cardiac device-related infective endocarditis.Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs · 2026
- Outcomes of Transcatheter Aortic Valve Replacement for Aortic Regurgitation in LVAD Patients: The JENA-VAD Study.Journal of the American College of Cardiology · 2026
- Development and preliminary characterization of a rapid porcine coronary vein graft failure modelFrontiers in Bioengineering and Biotechnology · 2026
- A BENCHTOP PLATFORM FOR EVALUATING SEPTOSTOMY IN AORTIC DISSECTION.Proceedings of the ... Design of Medical Devices Conference. Design of Medical Devices Conference · 2026
- Concomitant Wire-Loop Repositioning and Transapical Valve Replacement for a Migrated Sutureless Aortic Valve.JACC. Case reports · 2026