Default-threshold operating-point validation of a commercial chest radiograph AI system for selected CT-anchored thoracic findings: a bi-national multicenter retrospective study.
Yeliz Basar, Mustafa Ege Seker, Galina Ivanova Kirova-Nedyalkova, Elina Plamenova Milkovska, Boyan Emilov Manov, Atahan Tamturk, Ilke Tasci, Melih Karadag,
Department Radiology, Acibadem Maslak Hospital, Istanbul, Turkey. · Department of Radiology, University of Wisconsin-Madison, Madison, WI, USA.
Key points
From the paper's conclusion- This retrospective bi-national multicenter diagnostic accuracy study evaluated the locked default-threshold performance of a commercial chest radiograph AI system for selected thoracic findings using a CT-anchored, radiographic-detectability reference framework.
Abstract
This retrospective bi-national multicenter diagnostic accuracy study evaluated the locked default-threshold performance of a commercial chest radiograph AI system for selected thoracic findings using a CT-anchored, radiographic-detectability reference framework. Consecutive eligible adult patients who underwent frontal chest radiography and chest CT at four tertiary-care centers in Turkey and Bulgaria between June and December 2024 were included. Reference labels were assigned using temporally paired CT, with adjudication of whether CT-confirmed abnormalities had a corresponding radiographic manifestation on the paired chest radiograph. For potentially dynamic findings, the allowable CT-CXR interval was restricted to ≤ 2 days. The AI system was evaluated at the manufacturer's default threshold of 0.50. Diagnostic performance was summarized using prevalence, sensitivity, specificity, positive predictive value, negative predictive value, and false-positive burden with 95% confidence intervals. The study included 940 patients with paired CXR-CT examinations. Reference-positive prevalence ranged from 2.2% for pneumothorax to 21.0% for pleural effusion. Sensitivity was highest for fracture (91%) and pneumothorax (90%) and lowest for atelectasis (64%); specificity ranged from 82% for consolidation/opacity to 98% for fracture and pneumothorax. PPV ranged from 43% to 63%, indicating a non-trivial false-positive burden at the evaluated operating point. Because continuous probability scores, human-reader comparison, and workflow outcomes were unavailable, these findings should be interpreted as default-threshold technical validation and support further evaluation of the system as radiologist-supervised decision support with local performance monitoring, rather than standalone diagnosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- CT-Referenced Diagnostic Performance and Complementary Error Patterns of Artificial Intelligence Versus a Single Emergency Physician in Chest Radiograph Interpretation: A Retrospective Paired Diagnostic Accuracy Study.Tomography (Ann Arbor, Mich.) · 2026
- Diagnostic Performance of Different AI Tools for Chest Radiography: A CT-Anchored Per-Abnormality Analysis in a Tertiary Referral Center.Diagnostics (Basel, Switzerland) · 2026
- How threshold customisation affects the performance of a multiclass X-ray AI model for primary care triage: a retrospective studyBMJ Open · 2026
- Prospective Diagnostic Accuracy and Technical Feasibility of Artificial Intelligence-Assisted Rib Fracture Detection on Chest Radiographs: Observational StudyJMIR Medical Informatics · 2026
- Novel Artificial Intelligence Chest X-ray Diagnostics: A Quality Assessment of Their Agreement with Human Doctors in Clinical RoutineRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · 2026
- Diagnostic accuracy and clinical utility of AI-based chest radiograph assessment for osteoporosis: an updated systematic review and meta-analysis.Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA · 2026