Toward patient-centered blood pressure monitoring: first evaluation of a low-pressure and rapid oscillometric method for reducing cuff-related measurement burden.
Yuki Ota, Mitsuo Kuwabara, Kazuomi Kario
Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan. · Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan. kkario@jichi.ac.jp.
Key points
From the paper's conclusion- In conclusion, the novel low-pressure and rapid oscillometric method significantly reduced subjective and objective measures of cuff-related measurement burden, and measurements performed with the HEM-7290T were more frequently selected as involving the least cuff tightness than measurements performed using conventional methods.
Abstract
Out-of-office blood pressure (BP) monitoring is essential for hypertension management. Repeated cuff inflation may cause discomfort and reduce acceptance of out-of-office BP monitoring. To address this, we developed a novel low-pressure and rapid oscillometric method designed to reduce maximum cuff pressure (MaxCP) and measurement duration. In this participant-blinded crossover study, we evaluated cuff-related measurement burden using the Omron HEM-7290T, the first automated upper-arm BP monitor to use the novel method. The HEM-7290T was compared with an otherwise identical device using a conventional inflation-based method (Reference A; n = 110 participants) and with a conventional deflation-based oscillometric BP monitor (Reference B; n = 93 participants). The primary endpoint was the difference in visual analog scale (VAS) score for perceived cuff tightness. Compared with Reference A and Reference B, HEM-7290T reduced VAS scores by 8.6 points (95% CI, 6.5-10.8; p < 0.001) and 9.5 points (95% CI, 7.1-11.8; p < 0.001), respectively. Greater reductions in VAS scores were associated with more frequent selection of the HEM-7290T measurement as the one associated with the least perceived cuff tightness. Compared with Reference A and Reference B, respectively, MaxCP was reduced by 32.0 mmHg and 51.0 mmHg, while measurement duration was reduced by 6.8 s and 14.5 s. In conclusion, the novel low-pressure and rapid oscillometric method significantly reduced subjective and objective measures of cuff-related measurement burden, and measurements performed with the HEM-7290T were more frequently selected as involving the least cuff tightness than measurements performed using conventional methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Participant experience with the SpaceLabs 90227 ambulatory blood pressure monitor and SOMNOmedics ABPMpro device.Blood pressure monitoring · 2026
- An Estimation Framework for Cuff-based Blood Pressure Measurement: Analysis and Correction of Systematic Biases.Physiological measurement · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Cuffless Devices for the Measurement of Blood Pressure: A Scientific Statement From the American Heart AssociationHypertension · 2025
- Comparing noninvasive blood pressure monitoring technologies: the challenge of between-arm differences.British journal of anaesthesia · 2026