Toward patient-centered blood pressure monitoring: first evaluation of a low-pressure and rapid oscillometric method for reducing cuff-related measurement burden.

Yuki Ota, Mitsuo Kuwabara, Kazuomi Kario

Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan. · Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Tochigi, Japan. kkario@jichi.ac.jp.

Key points

From the paper's conclusion
  • In conclusion, the novel low-pressure and rapid oscillometric method significantly reduced subjective and objective measures of cuff-related measurement burden, and measurements performed with the HEM-7290T were more frequently selected as involving the least cuff tightness than measurements performed using conventional methods.

Abstract

Out-of-office blood pressure (BP) monitoring is essential for hypertension management. Repeated cuff inflation may cause discomfort and reduce acceptance of out-of-office BP monitoring. To address this, we developed a novel low-pressure and rapid oscillometric method designed to reduce maximum cuff pressure (MaxCP) and measurement duration. In this participant-blinded crossover study, we evaluated cuff-related measurement burden using the Omron HEM-7290T, the first automated upper-arm BP monitor to use the novel method. The HEM-7290T was compared with an otherwise identical device using a conventional inflation-based method (Reference A; n = 110 participants) and with a conventional deflation-based oscillometric BP monitor (Reference B; n = 93 participants). The primary endpoint was the difference in visual analog scale (VAS) score for perceived cuff tightness. Compared with Reference A and Reference B, HEM-7290T reduced VAS scores by 8.6 points (95% CI, 6.5-10.8; p < 0.001) and 9.5 points (95% CI, 7.1-11.8; p < 0.001), respectively. Greater reductions in VAS scores were associated with more frequent selection of the HEM-7290T measurement as the one associated with the least perceived cuff tightness. Compared with Reference A and Reference B, respectively, MaxCP was reduced by 32.0 mmHg and 51.0 mmHg, while measurement duration was reduced by 6.8 s and 14.5 s. In conclusion, the novel low-pressure and rapid oscillometric method significantly reduced subjective and objective measures of cuff-related measurement burden, and measurements performed with the HEM-7290T were more frequently selected as involving the least cuff tightness than measurements performed using conventional methods.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers