An Estimation Framework for Cuff-based Blood Pressure Measurement: Analysis and Correction of Systematic Biases.
Masoud Nateghi, Reza Sameni
Department of Biomedical Informatics, Emory University, 101 Woodruff Cir, Atlanta, Georgia, 30322, United States. · Emory University, 100 Woodruff Circle, Atlanta, Georgia, 30322, United States.
Key points
From the paper's conclusion- This study models and characterizes these biases and develops an estimation-theoretic framework to mitigate them.
Approach: Using the MIMIC database, we examine two major sources of BP measurement error: (1) systematic biases introduced by cuff-based occlusive measurement, and (2) respiratory-driven modulation of BP.
Abstract
Modern digital oscillometry and traditional auscultation-based blood pressure (BP) measurement share a similar cuff-based occlusive technique to sense BP: an external arm cuff is inflated above arterial pressure and then deflated while pressure changes due to arterial reopening and pulsation are detected and used to estimate BP. This process introduces inherent biases that can affect clinical assessments. This study models and characterizes these biases and develops an estimation-theoretic framework to mitigate them. Approach: Using the MIMIC database, we examine two major sources of BP measurement error: (1) systematic biases introduced by cuff-based occlusive measurement, and (2) respiratory-driven modulation of BP. Using a model of cuff-artery pressure interaction, we characterize errors in systolic BP (SBP) and diastolic BP (DBP) relative to a-line measurements, and analyze the association between respiration and short-term BP variability. We then evaluate least squares (LS), maximum likelihood (ML), and Bayesian estimators for single and multiple BP measurements using leave-one-subject-out cross-validation. Main Results: Simulated cuff-based measurements underestimated SBP by 9.7±7.8 mmHg and overestimated DBP by 6.6±4.8 mmHg on average. The proposed estimation frameworks substantially improved BP estimation errors in the simulated measurements. LS reduced estimation variability but retained the systematic bias, whereas ML corrected this bias, producing mean errors close to zero. Bayesian estimation further reduced mean absolute error (MAE) for most BP metrics. With three measurements, LS achieved MAEs of 9.7±5.7 mmHg and 6.6±3.6 mmHg for SBP and DBP, respectively, compared with 4.1±4.0 mmHg and 2.8±2.3 mmHg using ML. Bayesian estimation further reduced the MAEs to 3.6±4.4 mmHg and 1.9±2.1 mmHg, respectively. Significance: Cuff-based occlusive BP measurements are subject to inherent systematic biases. Modeling and correcting their systematic measurement errors can improve the accuracy and reliability of cuff-based noninvasive BP estimation, particularly when combined with repeated measurements and prior statistical information.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Membrane theory with multimodal evidence defines blood pressure measurementmedRxiv · 2026
- Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical TranslationarXiv · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Toward patient-centered blood pressure monitoring: first evaluation of a low-pressure and rapid oscillometric method for reducing cuff-related measurement burden.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- A dataset of paired auscultatory and oscillometric blood pressure measurements from routine clinical practice.Scientific data · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026