From the paper's conclusion- This study reports a non‐enzymatic electrochemical sensor based on a CoFe 2 O 4 /reduced graphene oxide (CoFe 2 O 4 /r‐GO) nanocomposite‐modified screen‐printed carbon electrode (CFO/r‐GO/SPCE).
Abstract
Uric acid (UA) is an endogenous antioxidant and a biomarker associated with physiological and pathological conditions. Abnormal UA levels are associated with hyperuricemia and the formation of monosodium urate (MSU) crystals, linked to gout and metabolic disorders. Therefore, simple, low‐cost, and reliable analytical platforms for UA detection are needed. This study reports a non‐enzymatic electrochemical sensor based on a CoFe 2 O 4 /reduced graphene oxide (CoFe 2 O 4 /r‐GO) nanocomposite‐modified screen‐printed carbon electrode (CFO/r‐GO/SPCE). Graphene oxide (GO) was synthesised by electrochemical exfoliation of graphite and subsequently reduced to obtain r‐GO, which acts as a conductive matrix and supports the dispersion of CoFe 2 O 4 nanoparticles. The electrochemical behaviour was investigated using cyclic voltammetry (CV) and differential pulse voltammetry (DPV). The results showed that the CoFe 2 O 4 /r‐GO modification enhances the charge‐transfer properties of the electrode compared to bare SPCE. The sensor exhibited a linear response to UA between 215–1725 µM, with limits of detection of 215.45 µM (CV) and 250.45 µM (DPV), and a sensitivity of 48.5 µA µM − 1 cm − 2 (CV). The effects of scan rate and pH indicated that UA oxidation follows with a diffusion‐controlled process involving proton transfer, suggesting a proton‐coupled electron‐transfer mechanism. Response surface methodology revealed significant interactions among experimental parameters affecting the electrochemical response.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dendritic Fe2O3-rGO nanoarchitectures for electrochemical uric acid sensing coupled with explainable machine learning-driven sensor analytics.Nanoscale advances · 2026
- Dispersive micro-solid phase extraction of uric acid from human serum using a CuFe2O4@Go-activated carbon hybrid nanocomposite prior to spectrophotometric determination.Analytical methods : advancing methods and applications · 2026
- AuNPs/reduced graphene oxide biointerface with immobilized uricase for mediator-free electrochemical uric acid biosensing.Analytical and bioanalytical chemistry · 2026
- Highly Sensitive Electrochemical Uric Acid Detection With Synergistic Integration of Exfoliated WS 2 Nanosheets and Ti 3 CN MXeneChemNanoMat · 2026
- Analytical quality by design for electrochemical nanostructured screen-printed sensor development: uric acid detection in saliva for diagnosis and monitoring of chronic diseases.Talanta · 2026
- Redox-Active Polyindole-Modified Electrode for Electrochemical Uric Acid Determination.ACS omega · 2026