ACS omega

Fe2O3-MgO-CuO@MWCNT Nanocomposite for Highly Sensitive Electrochemical Detection of l‑Cysteine at Low Potential in Biological Samples.

Danilo B Ribeiro, Carolina A Carvalho, Natilene M Brito, Flavio S Damos, Rita C S Luz

Universidade Federal do Maranhão, Avenida dos Portugueses, No 1966, Bacanga, São Luís, Maranhão CEP 65080-805, Brasil. · Instituto Federal de Educação, Ciência e Tecnologia do Maranhão, Avenida Getúlio Vargas, No 04, Monte Castelo, São Luís, Maranhão CEP 65030-005, Brasil.

Key points

From the paper's conclusion
  • The lowest detectable concentration of l-CYS with the Fe2O3–MgO–CuO@MWCNT/GCE electrode was 0.5 μmol L–1.

Abstract

A Fe2O3-MgO-CuO composite integrated with multiwalled carbon nanotubes (MWCNTs) was employed to develop a platform for the indirect electrochemical detection of l-cysteine (l-CYS). The material was characterized by XRD, FTIR, SEM, EDS, and EIS, confirming the formation of a hybrid nanostructure. After optimization of the experimental parameters, the modified glassy carbon electrode exhibited a well-defined cathodic response at -0.25 V vs Ag/AgCl (saturated KCl). Although l-CYS undergoes electrochemical oxidation at positive potentials (approximately +0.25 V), quantitative analysis was based on the effect of the presence of l-CYS on the cathodic response of the modified electrode. The determination of the analyte was performed using linear sweep voltammetry, yielding two linear response ranges of 0.5-100 μmol L-1 (R2 = 0.992) and 100-1000 μmol L-1 (R2 = 0.996). The lowest detectable concentration of l-CYS with the Fe2O3-MgO-CuO@MWCNT/GCE electrode was 0.5 μmol L-1. The sensor exhibited high selectivity, excellent repeatability (RSD = 1.58%), and reproducibility (RSD = 1.82%). The sensor was evaluated for l-CYS determination in urine and breast milk samples, providing satisfactory recovery values.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers