A Heating-Assisted Integrated Purification-Enrichment-Derivatization Paper Chip for Rapid MS/MS Quantification of Serum Steroid Hormones.
Can Tan, Zhangwu Chen, Jingxiong Lin, Yue Zhang, Xunliang Su, Xuan Yu, Zhenqing Cai, Le Wang,
School of Biomedical Engineering, Shenzhen Campus of Sun Yat-sen University, Shenzhen, Guangdong518107, P. R. China. · Shenzhen Bioeasy Biotechnology Co., Ltd., Shenzhen, Guangdong518126, P. R. China.
Key points
From the paper's conclusion- Coupled with tandem mass spectrometry, the iPED chip-MS/MS method enabled rapid simultaneous quantification of testosterone (T), androstenedione (4-AD), dehydroepiandrosterone (DHEA), and 17α-hydroxyprogesterone (17-OHP) in serum within 20 min.
Abstract
Rapid and reliable simultaneous quantification of multiple steroid hormones is important for the diagnosis and monitoring of endocrine disorders such as polyendocrine metabolic ovarian syndrome (PMOS), formerly known as polycystic ovary syndrome (PCOS). Existing methods are susceptible to ion suppression in complex serum matrices and therefore usually require laborious sample preparation and time-consuming chromatographic separation, limiting their application in high-throughput screening and rapid testing. In this study, we developed an integrated purification-enrichment-derivatization paper chip, termed iPED chip, which integrates diatomaceous earth-assisted online cleanup at the tail end, thermal enrichment at the paper tip, and in situ hydroxylamine derivatization. A programmable induction heating device was also developed to provide a localized, controllable, and noncontact thermal field for the paper reaction zone, thereby promoting analyte enrichment and reaction. Coupled with tandem mass spectrometry, the iPED chip-MS/MS method enabled rapid simultaneous quantification of testosterone (T), androstenedione (4-AD), dehydroepiandrosterone (DHEA), and 17α-hydroxyprogesterone (17-OHP) in serum within 20 min. Method characterization demonstrated effective enrichment of the target hormones, broad linearity over 0.078-40 μg/L in spiked calf serum, low limits of detection of 0.042-0.093 μg/L, recoveries of 95-106%, and RSDs below 10%. In spiked hormone-depleted human serum samples at clinically relevant levels, the results agreed well with those obtained by conventional LC-MS/MS, supporting the reliability of iPED chip-MS/MS and its potential for rapid clinical testing and large-scale biofluid screening.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-338 Magnetic Bead–Based Sample Preparation for LC–MS/MS Quantification of Steroid Hormones Using PEP Polymer beadsClinical Chemistry · 2026
- Development and validation of a comprehensive mass spectrometry method for the determination of steroids in multi-matrixJournal of Chromatography Open · 2026
- Optimisation and Validation of a UHPLC-MS/MS Assay Using Timed Post-Column Ammonium Fluoride Infusion for Sensitive Serum Steroids QuantitationbioRxiv · 2026
- Detection of estrogens using 2-furoyl chloride derivatization coupled with ultra-high-performance liquid chromatography-quadrupole time-of-flight mass spectrometry.Journal of chromatography. A · 2026
- Determination of hydrocortisone and cortisone in artificial saliva by spray assisted fine droplet formation liquid phase microextraction coupled to LC–MS/MSScientific Reports · 2026
- Automated enzymatic hydrolysis-magnetic-bead-assisted sequential extraction workflow for LC-MS/MS profiling of urinary aldosterone-pathway steroids.Talanta · 2026