Enhanced Microplasma Excitation Source Using Fiber-Tip Discharge Clusters with Low Breakdown Voltage for Miniaturized Optical Emission Spectroscopy.
Yan Zhang, Yan Zhang, Yong-Liang Yu, Shuang Liu
Department of Forensic Analytical Toxicology, School of Forensic Medicine, China Medical University, Shenyang110122, China. · Research Center for Analytical Sciences, Department of Chemistry, College of Sciences, Northeastern University, Shenyang110819, China.
Key points
From the paper's conclusion- This work provides an electrode-interface engineering strategy for rapid multielement screening of trace samples and portable heavy metal analysis.
Abstract
Microplasma based optical emission spectrometry (OES) holds great promise for field elemental analysis, yet its practical application is hindered by limited excitation capability and high breakdown voltage constrained by millimeter-scale discharge geometries. Herein, we report an enhanced fiber-tip discharge microplasma excitation source constructed from carbon felt electrodes, enabling simultaneous multielement analysis of microliter-scale samples. During discharge, sharp tips are generated at the ends of micrometer-scale carbon fibers at the carbon felt interface, providing abundant high-field sites and promoting initial ionization. The resulting multitip configuration provides densely distributed fiber-tip discharge sites, reducing the breakdown voltage and improving local energy deposition for analyte excitation. By integrating sample loading, drying of the preloaded carbon felt electrode, and in situ excitation, the system enables efficient analysis of 50 μL solution samples. Under the same operating conditions, the carbon felt electrode pair provided clearly detectable multielement emission signals that were difficult to obtain using conventional point discharge electrode configurations. This strategy was applied to the simultaneous determination of Cr, Mn, Cd, and Pb, covering hydride-forming and nonhydride-forming elements, with limits of detection (LODs) of 0.3, 0.1, 0.2, and 0.2 μg L-1, respectively, and relative standard deviations (RSDs) below 5% (n = 11). The accuracy and applicability were validated through analysis of certified reference materials and real samples. With its compact configuration, low sample requirement, and reliable analytical performance, this method shows promise for on-site elemental analysis. This work provides an electrode-interface engineering strategy for rapid multielement screening of trace samples and portable heavy metal analysis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Enhanced Electrolyte Atmospheric Liquid Discharge via Three-Dimensional Synergistic Effect for Trace Multielement Optical Emission SpectrometryAnalytical Chemistry · 2026
- T-shaped confined microplasma system for rapid colorimetric detection of Cr(VI).Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- Microplasma glow discharge chip for open-air atomic emission spectroscopyJournal of Micromechanics and Microengineering · 2026
- Signal enhancement in laser-induced breakdown spectroscopy for improved sensitivity and detection limits: Mechanisms and prospects.The Review of scientific instruments · 2026
- A compact short-path electrothermal vaporization-atmospheric pressure glow discharge atomic emission spectrometry device for the determination of cadmium and lead.Analytical methods : advancing methods and applications · 2026
- Augmenting ICP-OES Sensitivity for Trace Element Analysis in Crude Oil via High-Efficiency Micro-Ultrasonic Nebulization.Analytical chemistry · 2026