Analytical ChemistryChemistry

Measurement of the Isotopic Composition of Uranium in Geological Samples by MC-ICP-MS Using a TOPO-Based Extraction Chromatographic Protocol

Jia-Kai Guo, Qiaohui Zhong, Mei-ju Li, Gao-bin Chu, Zhao‐Yang Wang, Wei-hai Xu, Neng-ping Shen, Jie Li

Chinese Academy of Sciences · Guangzhou Electronic Technology (China)

Key points

From the paper's conclusion
  • Overall, this protocol enables the quantitative recovery of high-purity U and high-precision U isotope determination, providing an alternative to the double spike technique and promoting its wider application in geochemistry.

Abstract

The precise determination of uranium (U) isotope ratios in geological samples is crucial for tracing redox processes and reconstructing paleoenvironments. However, the widespread application of U isotopes is often limited by the difficulty of obtaining a 233 U– 236 U double spike for correction of instrumental isotopic fractionation. In this study, a single-column purification procedure was used to separate U from geological samples using a self-made tri- n -octylphosphine oxide (TOPO) extraction resin. This procedure takes advantage of the high adsorption selectivity of U on the TOPO resin, achieving low procedural blanks, quantitative U recovery (100.1 ± 1.6%), and effective separation from major and matrix elements (e.g., Ti, Fe, Zn, Pb, Zr, and Th). A combined standard-sample bracketing and internal normalization (C-SSBIN) method, using NIST SRM 981 Pb as the normalization element, was applied to correct instrumental isotopic fractionation during U isotope measurements by MC-ICP-MS. Repeated measurements of the standard solutions GBW04428 and HPS U yielded a long-term external reproducibility (2SD) for δ 238 U of ±0.064‰, comparable to that achieved by double spike methods in previous studies. Repeated analyses of a suite of geological and environmental reference materials (BCR-2, GSR-1, GSR-3, SGR-1b, GSR-12, NOD-P-1, JCP-1, and seawater) yielded δ 238 U values in excellent agreement with published data, demonstrating the accuracy and robustness of the protocol. Overall, this protocol enables the quantitative recovery of high-purity U and high-precision U isotope determination, providing an alternative to the double spike technique and promoting its wider application in geochemistry.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Measurement of the Isotopic Composition of Uranium in Geological Samples by MC-ICP-MS Using a TOPO-Based Extraction Chromatographic Protocol — Science Skim