Electrospray-Orbitrap Analysis of Nitrate and Sulfate Isotopologues in Natural Samples: Performance and Constraints.
Jack Saville, Julien Witwicky, Elsa Gautier, Alexis Lamothe, Nils Johannes Kuhlbusch, Becky Alexander, Andrew Schauer, Andrea Burke,
Univ. Grenoble Alpes, CNRS, IRD, Grenoble-INP, INRAE, IGE, Grenoble38000, France. · University of Münster, Münster48149, Germany.
Key points
From the paper's conclusion- We resolve relative nitrate clumped isotope values in ice core samples (Δ15N18O-NO3-: ±2.3‰; Δ18O18O-NO3-: ±3.3‰) yet observe no correlations with bulk nitrate isotopes.
Abstract
Traditional isotope ratio mass spectrometers are not optimally designed to analyze nonvolatile molecular solutes such as oxyanions or organics. Necessary sample conversion steps, often demanding micromole-scale samples, break chemical bonds and limit oxyanion analyses to elemental isotope ratios. Recent studies enthusiastically promote soft electrospray ionization coupled to high-resolution Orbitrap mass spectrometry (ESI-Orbitrap-MS) for intact-molecule isotopologue analysis. This combination uses widely accessible instrumentation to provide simultaneous multi-isotopologue and clumped isotope analyses using tens of nanomoles of dissolved sample. However, its isotopic capabilities have been evaluated primarily through technical assessments rather than on natural samples, despite the need for matrix sensitivity assessments and robust isotopic calibrations. Here, we establish a framework for natural-sample ESI-Orbitrap-MS isotopologue analysis of nitrate and sulfate, addressing matrix effects, isotopologue calibration, and instrumental stability. We demonstrate strong sensitivity to contaminants at concentrations 1000 times below target analytes and strong temperature-correlated drift over >30 min, highlighting the enhanced need for rigorous sample purification to suppress matrix effects, sample-reference bracketing to overcome drift, and isotopologue calibration to reduce biases and robustly quantify uncertainties. We achieve 0.4-1.3‰ precision and accurate measurements of δ15N-NO3-, δ18O-NO3-, Δ17O-NO3-, δ34S-SO42-, and Δ17O-SO42- relative to traditional techniques, consuming only 20 nmol nitrate or 10 nmol sulfate. We resolve relative nitrate clumped isotope values in ice core samples (Δ15N18O-NO3-: ±2.3‰; Δ18O18O-NO3-: ±3.3‰) yet observe no correlations with bulk nitrate isotopes. These results establish practical guidelines for applying ESI-Orbitrap-MS to isotopologue and clumped isotope analysis of trace analytes in challenging sample matrices under real-world conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Lessons Learned in Orbitrap MS-Based Isotope Ratio Analysis of Organic Acid MixturesAnalytical Chemistry · 2026
- Oxygen Isotope Analyses of Phosphate and Organophosphorus Compounds by Electrospray Ionization Orbitrap Mass SpectrometryAnalytical Chemistry · 2025
- Online Two-Dimensional Ion Chromatography Coupled with Orbitrap MS for Measuring the Triple Isotopic Composition (δ15N, δ18O, Δ17O) of Nitrate.Analytical chemistry · 2026
- Intramolecularly Labeled Reference Standards of Sulfamethoxazole for Fragment-Specific Isotope Analysis by Electrospray Ionization Orbitrap Mass SpectrometryJournal of the American Society for Mass Spectrometry · 2026
- δ13C Determination of Benzoic Acid by High-Resolution Orbitrap MS: How Sample Introduction Strategy Governs Accuracy and Precision.Journal of the American Society for Mass Spectrometry · 2026
- IsoSwap: A Toeplitz Operator Framework for Isotope Transformations in Mass Spectrometry.Analytical chemistry · 2026