Clinical application of MR-Only in cyberknife stereotactic radiotherapy for brain metastases: a feasibility study.
Yajuan Wang, Zhongming Yin, LiMing Sheng, Pengfei Zhang, Ziquan Wei, Lecheng Jia, Jiping Liu, Binbing Wang,
Zhejiang Cancer Hospital, Hangzhou, Zhejiang, 310022, China. · Radiotherapy Business Unit, Shanghai United Imaging Healthcare Co., Ltd, Shanghai, China.
Key points
From the paper's conclusion- The proposed model can generate accurate sCT for brain metastases, and the feasibility of the MR-only in CyberKnife SRT has been verified.
Abstract
BACKGROUND: Due to the limited number of studies on MR-only stereotactic radiotherapy (SRT) for brain metastases using CyberKnife in the literature, this study systematically evaluates the dose calculations and image guidance accuracy of synthetic CT (sCT) and sCT-derived digital reconstructed radiographs (sDRRs), to assess the feasibility of clinical MR-only workflow implementation in CyberKnife SRT for brain metastases. METHODS: We developed a machine learning model trained on T1-weighted MR and corresponding CT images from 50 patients with brain metastases who had previously undergone CyberKnife treatment, with images from an additional 18 patients used to test the feasibility of the MR-only workflow. The Mean Absolute Error (MAE), Structural Similarity Index (SSIM), Peak Signal-to-Noise Ratio (PSNR) and the Dice Similarity Coefficient (DSC) were used to evaluate the image quality of sCT and sDRR. For dosimetric evaluation, the sCT-based and transferred plans (namely, the original planning parameters were transferred to the sCT images for dose recalculation) was compared with the plans based on the real CT images (rCT) using DVH metrics. For image-guided positioning accuracy, the 2D errors generated by registering DRR with the digital radiography images (DRs) were converted into translational and rotational couch shifts using a geometric back-projection method, thereby assessing the consistency of positioning accuracy between the sDRRs with the rCT-derived DRRs (rDRRs). RESULTS: The sCT images had a mean MAE of 10.08 ± 2.14 HU, a SSIM of 0.98 ± 0.01, and a PSNR of 34.73 ± 1.27 dB, and the overall sDRRs maintained high image quality (MAE: 37.36 ± 9.31 HU; SSIM: 0.95 ± 0.03; PSNR: 31.41 ± 7.14 dB). Large OARs (brainstem/eyeballs) had high DSC (> 0.80), while small OARs (lenses/optic pathway) had lower DSC. The dosimetric parameters of PTV and OARs under both planning strategies showed no statistically significant difference from the rCT-based plans (p > 0.05). When using sDRRs for image guidance, the translational errors were submillimeter, and the rotational errors had a wider distribution but mean values within robotic correction range, there was no statistical difference compared with that of the rDRRs (p > 0.05). CONCLUSIONS: The proposed model can generate accurate sCT for brain metastases, and the feasibility of the MR-only in CyberKnife SRT has been verified. TRIAL REGISTRATION: This study is retrospectively registered in the Clinical trial Center of Zhejiang Cancer Hospital in Jan 1th, 2024 (No. MR-33-24-003134).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical implementation of deep learning-based synthetic CT for MRI-only volumetric modulated arc therapy in head and neck and pelvic cancer patientsRadiation Oncology · 2025
- Impact of dose calculation algorithm on magnetic resonance imaging-only radiotherapy for prostate and glioma.Physics and imaging in radiation oncology · 2026
- Application of synthetic computed tomography based on a deep learning model in pelvic radiotherapy planning: A feasibility study.Journal of applied clinical medical physics · 2026
- Integrating functional brain regions into treatment optimization: a multi-institutional dosimetric study in CyberKnife SRT for brain metastases.Radiation oncology (London, England) · 2026
- MRI-only workflow for preclinical ion beam therapy: Deep-learning-based synthetic CT generation from ultra-high-field small animal MRI.Zeitschrift fur medizinische Physik · 2026
- Magnetic Resonance Only Planning Validation Using Synthetic CT by MVisionVCU Scholars Compass (Virginia Commonwealth University) · 2026